C 150 (E2) 06/2011
page 64/78
•
After a manual abortion, the C 150 reverts to its standard operational status. The indication “ - - - “ in
the lower display shows that the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
•
If necessary, repeat the sterilization.
•
When the inner chamber has cooled down to a value below 60 °C /
140°F
, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C /
140°F
.
•
Turn on the unit (chap. 5).
15.5.2 Aborting the hot-air sterilization by opening the outer door
For safety reasons, the sterilization cycle is automatically aborted if you open the outer door of the C 150.
•
Do not open the glass door and close the outer door immediately.
•
After an abortion caused by opening the outer door, the C 150 reverts to its standard operational
status. The indication “ - - - “ in the lower display shows that the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
•
If necessary, repeat the sterilization.
•
When the inner chamber has cooled down to a value below 60 °C /
140°F
, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C /
140°F
.
•
Turn on the unit (chap. 5).
15.5.3 Aborting the hot-air sterilization using the main power switch
For safety reasons, the sterilization cycle is aborted if the C 150 is turned off or if there is a power failure.
Normally you should not terminate sterilization in this way.
•
Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C /
98.6°F
.
•
After turning on again the C 150, it reverts to its standard operational status. The indication “ - - - “ in
the lower display shows that the CO
2
sensor is disconnected.
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing “EXIT”.
•
If necessary, repeat the sterilization.
•
When the inner chamber has cooled down to a value below 60 °C /
140°F
, turn off the unit and plug in
the CO
2
sensor.
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C /
140°F
.
•
Turn on the unit (chap. 5).