![Binder 9020-0213 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/binder/9020-0213/9020-0213_translation-of-the-original-operating-manual_2752637048.webp)
UF V (E2+E2.1) 03/2013
page 48/112
5.5
Behavior after turning on the unit
During the equilibration phase of 9 hours after turning on the unit, undefined temperature conditions oc-
cur within the unit. During this phase, do not place any sample materials in the unit
CAUTION
Danger of destruction of samples.
Load the freezer only after equilibration of temperature.
Following turning on the unit with the key switch (6), there is a subsequent brief startup phase in which
the display items at the edges of the controller display light up successively.
After a few seconds, the display shows the current interior temperature of the freezer. The lit LED at the
right of the display shows that the temperature is displayed in °C.
Figure 29: Normal display of controller after turning on the
freezer, showing the current temperature value: 15 °C (ex-
ample)
Set the controller to the desired temperature set point (chap. 6) used to operate the freezer.
As long as there is a difference between the actual and set value shown in the display, the
intended operation of the unit will not be ensured.
Содержание 9020-0213
Страница 6: ...UF V E2 E2 1 03 2013 page 6 112...
Страница 7: ...UF V E2 E2 1 03 2013 page 7 112 Product registration...
Страница 104: ...UF V E2 E2 1 03 2013 page 104 112 18 7 Dimensions UF V 500 E2 xxx Indications in mm...
Страница 105: ...UF V E2 E2 1 03 2013 page 105 112 18 8 Dimensions UF V 500 E2 1 xxx Indications in mm...
Страница 106: ...UF V E2 E2 1 03 2013 page 106 112 18 9 Dimensions UF V 700 E2 1 xxx Indications in mm...
Страница 113: ......