UF V (E2+E2.1) 03/2013
page 42/112
5.
Start up
Following transport, wait at least 8 hours until start-up.
Check that the interior of the freezer is empty. Prior to starting a new freezer
or if you do not know what the freezer was last used for, for hygiene purpos-
es you should clean and disinfect the interior (chap.14.3).
After connecting the supply lines, turn on the unit by the key switch (6) set-
ting it to position 1).
Figure 24: Key switch (position ON)
5.1
Preset factory parameters
The unit is supplied with the following preset parameters:
Temperature set point
-80 °C /
-112 °F
Safety controller
Set point type “Limit”
-60 °C /
-76 °F
The maximum permitted deviation from the temperature set point for
tolerance range alarm
6 K
Delay time tolerance range alarm temperature after closing the door
60 min
Password to lock / unlock the HAND operating functions
0 (no locking)
Password to access operating mode USER
1
5.2
Operating the automatic door mechanism
Opening the door:
Firmly press the one-touch pushbutton “OPEN”. You can now open
the door.
Closing the door:
Firmly press on the door for at least 2 seconds until the automatic
door mechanism is activated and closes the door tightly.
Figure 25: Pushbutton “OPEN” to
open the door
After turning on the freezer, you can open the door after 30 seconds only.
Following closing the door, you can open the door again only after approx. 30 seconds due to
necessary venting.
Содержание 9020-0213
Страница 6: ...UF V E2 E2 1 03 2013 page 6 112...
Страница 7: ...UF V E2 E2 1 03 2013 page 7 112 Product registration...
Страница 104: ...UF V E2 E2 1 03 2013 page 104 112 18 7 Dimensions UF V 500 E2 xxx Indications in mm...
Страница 105: ...UF V E2 E2 1 03 2013 page 105 112 18 8 Dimensions UF V 500 E2 1 xxx Indications in mm...
Страница 106: ...UF V E2 E2 1 03 2013 page 106 112 18 9 Dimensions UF V 700 E2 1 xxx Indications in mm...
Страница 113: ......