MK (E2.1) 10/2014
page 63/73
18.4 Technical Data MK 53 (E2.1)
Exterior dimensions
Width (including access port 80 mm with plug)
mm /
inch
745 /
29.33
Height (feet included)
mm /
inch
1245 /
49.02
Depth (including 55 mm controller and door handle)
mm /
inch
795 /
31.30
Wall clearance sides
mm /
inch
160 /
6.30
Wall clearance rear
mm /
inch
100 /
3.94
Window width
mm /
inch
280 /
11.02
Window height
mm /
inch
280 /
11.02
Interior dimensions
Width
mm /
inch
402 /
15.83
Height
mm /
inch
402 /
15.83
Depth
mm /
inch
330 /
12.99
Interior volume
l /
cu.ft.
53 /
1.9
Number of racks ( standard / max.)
1/5
Load per rack
Kg /
lbs.
15 /
33
Permitted total load
Kg /
lbs.
40 /
88
Temperature data
Temperature range
°C /
°F
-40 to +180
Temperature uniformity (variation)
± K
0.4 to 2.0
Temperature fluctuation
± K
0.1 to 0.5
Recovery time after door was
opened for 30 sec 1)
at -10 °C
minutes
5
at + 70 °C
minutes
1
at +150 °C
minutes
5
Average heating up time from -40 °C to 180 °C
K/min.
4.6
Average cooling down time from 180 °C to -40 °C
K/min.
4.1
Max. heat compensation
W
500
Electrical data
IP-system of protection acc. to EN 60529
IP
20
Nominal voltage (+10%) 50 Hz
V
230 1N~
Power plug
shockproof plug
Over-voltage category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Nominal power
kW
2.60
Energy consumption 2) at 20 °C /
68 °F
Wh/h
1020
Further information
Weight (empty)
Kg /
lbs.
150 /
331
Filling weight of refrigerant
Kg /
lbs.
0.85 /
1.87
Noise level
approx. dB(A)
59
1) to 98% of the set value
2) These values can be used for dimensioning air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C /
77 °F
and a power supply voltage fluctuation of ±10. The temperature data is determined in ac-
cordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances
of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.