FDL (E2) 03/2011
page 38/68
8.4
Deleting a program section
A program section is deleted from the program by setting the section duration to Zero.
Normal display
Press down key
for 5 sec.
Display 1 shows
e.g. 0000
Display 2 shows PROGRAM EDITOR
(you are in the program editor)
Press program key
Display 1 shows
0000
Display 2 shows
UserCod? 0000
(enter user code)
Enter the user code using arrow keys
e.g.
0001
(basic setting, adjustable in the user level,
chap. 10).
Value is shown in both displays.
Automatically forward after 2 sec
Display 1 shows
e.g. 01
(program P01 selected)
--- : --- PRG.
(program can be selected)
Display 2 shows
alternating
CONTINUE X/W
(information: to 1
st
program section with X/W)
Select the desired program, e.g. P01, using
arrow keys
Value is shown in display1.
Press key
In the selected program P01 or P02, program sections can be selected:
Display 1 shows
e.g. 01
(actual selection of the section: S01)
P01: --- SEC.
(program section can be selected)
Display 2 shows
alternating
CONTINUE X/W
(information: set-point entry with X/W)
Select desired section, e.g. S03, using arrow
keys
(omitted if section S01 shall be deleted).
Press key
Display 1 shows
e.g. 90.0
C
(actual temperature set-point)
S03:TEMP 90.0
(variable: temperature)
Display 2 shows
alternating
CONTINUE X/W
(information: continue with X/W)
No entry
Press key
Display 1 shows
e.g. 01.00
(actual section length)
S03:TIME 01:00
(variable: section length)
Display 2 shows
alternating
CONTINUE X/W
(information: continue with X/W)
Enter set-point
Zero
for section length in
hh:mm of S03 using arrow keys
Value is shown in display 2 or in both displays
(display depends on maximum time setting in
the user level chap. 10)
Display 1 shows
e.g. 00.00
(actual section length)
S03:TIME 00:00
(variable: section length)
Display 2 shows
alternating
DELETE SEC. X/W
(information: delete section with X/W)
Press key