FED / FED-UL (E2) 01/2022
page 35/60
8. Options
8.1 APT-COM™ 4 Multi Management Software (option)
The chamber is regularly equipped with a serial interface RS 422 that can connect the BINDER APT-
COM™ 4 Multi Management Software. The connection to a computer is established using the chamber’s
interface via an interface converter.
Make sure that the interface mode is correctly set to
“0” = “Modbus”
in the user level (chap.
The actual temperature, and fan speed values are given at adjustable intervals. Programming can be
performed graphically via PC. Up to 100 chambers can be cross linked. For further information, refer to the
APT-COM™ 4 operating manual.
Pin allocation of the RS 422 interface: pin 2: RxD (+)
pin 3: TxD (+)
pin 4: RxD (-)
pin 5: TxD (-)
pin 7: Ground
If several chambers are to be recorded via a PC, each one must be allocated a unique
address. Addressing is performed via the chamber controller (see chap. 6.4.3).
8.2 HEPA fresh air filter (option)
With this option, the introduced fresh air is cleaned by means of a high efficiency submicron particulate air
filter type HEPA class H 14 (acc. to DIN EN 1822:2009). Replace the filter insert, if necessary, by removing
the metal cover of the filter at the left side of the chamber (Art. No. 6014-0003).
8.3 Analog output for temperature (option)
With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output
permits transmitting data to external data registration systems or devices.
The connection is carried out as a DIN socket at the rear of the chamber as follows:
ANALOG OUTPUT 4-20 mA DC
PIN 1: temperature –
PIN 2: tempe
Temperature range:
0 °C to +300 °C
A suitable DIN plug is enclosed.
Figure 8: Pin allocation of DIN socket for option analogue outputs