BD / ED / FD (E2) 01/2022
page 22/66
6. Operating the controller
Controller setting is identical with all three chambers BD, ED, and FD. The temperature controllers only
differ in their temperature range (BD: up to 100 °C /
212 °F
, ED/FD: up to 300 °C /
572 °F
) and the displayed
accuracy (BD: a tenth of a degree, ED/FD: one degree).
6.1 Display / entry of temperature set-point (without ramp function)
The chamber is operating, the controller is in normal display (actual value
display). The actual temperature value (example: 22 °C) is displayed:
1.
Press
button
The display shows alternately “
SP
” and the previous
temperature set-point
(example: 60 °C):
2.
With the
buttons enter a set-point value between 0 and 300.
The desired temperature set-point can be selected in a temperature range from 5 °C /
9 °F
above room temperature up to 100 °C /
212 °F
(BD) or 300 °C /
572 °F
(ED/FD).
3.
Wait 2
seconds until the entered temperature value
is
taken over
(display flashing once).
4.
Press
button to return to normal display (actual value display) (automatically after approx. 30
seconds).
Adjust the safety device following any changes of the set-point (chap. 7).
6.2 Display / entry of temperature set-point (with selected temperature ramp)
If previously a temperature ramp value has been selected (chap. 6.4.2):
Press button
in normal display / actual value display during ramp operation to have displayed the
actual temperature ramp set-point, which changes according to the selected gradient, in addition to the
entered target set-point for temperature.
The chamber is operating, the controller is in normal display (actual value
display). The
actual temperature value
(example: 22 °C) is displayed:
1.
Press
button
The display shows alternately “
SPr
” and the
actual temperature ramp set-point
changing according
to the selected gradient (example: 42 °C):
This ramp set-point is displayed only, it is not adjustable.
2.
Press
button
Содержание 9010-0073
Страница 54: ...BD ED FD E2 01 2022 page 54 66...
Страница 55: ...BD ED FD E2 01 2022 page 55 66 13 2 EU Declaration of Conformity for ED...
Страница 56: ...BD ED FD E2 01 2022 page 56 66...
Страница 57: ...BD ED FD E2 01 2022 page 57 66 13 3 EU Declaration of Conformity for FD...
Страница 58: ...BD ED FD E2 01 2022 page 58 66...
Страница 60: ...BD ED FD E2 01 2022 page 60 66...
Страница 67: ......