5.4 Operating modes
The terminal operating mode is set during installation.
On switching on the terminal, the selected operating mode is displayed.
5.4.1 Standard operation
In standard operating mode, the terminal can acquire the gross (or net)
weight on the scale and display it along with the weighing symbols. In
this case, the possible display formats are
15mm digits,
30mm
digits
and
Selection
(see par. 5.2.4 on page 1-42 ).
5.4.2 Sum weighing operation
The sum weighing operating mode allows you to perform weighing
operations in sequence without unloading the scale, zero-setting the net
weight after each operation. To move from one weighing operation to the
next, press
when the weight is stable.
On completion of this operation:
✔
thenet weight is zero-set and the weight currently on the scale is
taken as the tare;
✔
a string is sent to the serial port or the data is printed (if a printer is
connected).
The display formats available in this operating mode are
15mm
digits,
30mm digits
and
Selection
(see par. 5.2.4 on page
5.4.3 Loading extraction operation
This operating mode, when selected and associated with the I/O
devices, allows you to perform simple dosing operations with a number
of different components.
The I/O are selected during the installation phase.
Comply with the instructions in par. 5.2.8 on page 1-46 and enter the
Set Point 01 and Set Point 02 values. Set Point 01 defines the final
weight value to extract, Set Point 02 gives the weight value that,
subtracted from the Set Point 01 value, marks the change from fast
D410