background image

 
EF, AF 

 

 

LF01 

1.0  I

NTRODUCTION

 

1.1  Models Covered by this Manual 

This manual covers current models EF and AF. Please note that changes were made to the range 
from late 2001; these changes affect both the case and the internal design. Models made after the 
change are suffixed with the letter B on the rating label. 

If a manual for the earlier design is 

required, please contact Lenton. 

Model EF 11/8B may be set to operate on any voltage in the range 200-240V  (or 100-120V to 
order) by correct setting of the power limit parameter in the controller. For model AF 11/6B the 
equivalent range is 220-240V. 

Because of the model changes it is very important when ordering spares to state the furnace serial 
number or otherwise to identify the model correctly. 

1.2  Switches and Lights 

 

 

 

 

 

 

 

 

1.3  Warning Symbols 

 

 

 

 

 

 

 

 

 

 

 

 

 

Supply Light: when the furnace is connected to the electrical supply the 
light in the adjacent switch glows 

 

Heat Light: the adjacent light glows or flashes to indicate that power is 
being supplied to the elements 

 

DANGER of electrical shock– read any warning printed by this symbol. 

DANGER – hot surface. Read any warning printed by this symbol. 

WARNING: all surfaces of a furnace may be hot. 

 

DANGER – read any warning printed by this symbol. 

 

Содержание Lenton AF 11/6

Страница 1: ...onnected to the electricity supply Manuals are supplied separately for the furnace controller and overtemperature controller when fitted Please read the controller manuals before operating the furnace...

Страница 2: ...mit parameter in the controller For model AF 11 6B the equivalent range is 220 240V Because of the model changes it is very important when ordering spares to state the furnace serial number or otherwi...

Страница 3: ...the furnace on an inflammable surface Ensure that there is free space around the furnace Do not obstruct any of the vents in the case they are needed to keep the controls and the case exterior cool En...

Страница 4: ...TION DETAILS supply type Supply Terminal label Cable colour Live Neutral Reversible or Live Live 1 phase L Brown To live to either power conductor N Blue To neutral to the other power conductor PE Gre...

Страница 5: ...e temperature slowly over the last 300 C allowing 10 minutes soak at 100 C intervals When heating large objects in particular poor conductors avoid shielding the thermocouple from the heating elements...

Страница 6: ...he original setting before altering the power limit Example It is desired to run the furnace at 300 C The normal control settings can be expected to cause excessive overshoot as the furnace reaches te...

Страница 7: ...t should be isolated when the door is opened 4 2 Calibration After prolonged use the controller and or thermocouple could require recalibration This would be important for processes which require accu...

Страница 8: ...n Bulletin No 11 and the UK Health and Safety Executive Guidance Note EH46 We can provide further information on request Alternatively our service division can quote for any repairs to be carried out...

Страница 9: ...the chamber back blanket packing around the back and the chamber base board Disconnect the furnace from the supply and remove the back panel Remove the hearth tile if necessary See section 5 6 above t...

Страница 10: ...tain any insulation sleeves to refit later To remove the chamber box slacken the 4 bolts on the underside then slide the chamber box out the back of the case Be careful the chamber is heavy Remove the...

Страница 11: ...e are no lights glowing on the controller The SUPPLY light is ON The controller may be faulty or not receiving a supply due to a faulty switch or a wiring fault The SUPPLY light is OFF Check the suppl...

Страница 12: ...TE models before mid 2002 in some models two door switches were fitted directly in the element circuit instead of a single switch and contactor For models up to 16A the EMC filter and fuses are integr...

Страница 13: ...uses Required if no supply cable fitted Recommended if cable fitted See rating label or table below for recommended fuse rating Model phases Volts Supply Fuse Rating EF 11 8B 1 phase 200 240V 10A EF 1...

Страница 14: ...EF AF 14 LF01...

Страница 15: ...EF AF LF01 15...

Страница 16: ...erves the right to change specifications without notice 9 1 Models Covered by this Manual MODEL Max Temp Max Power Chamber Size mm Approx Capacity Net Weight C kW H W D l kg Chamber furnace heated by...

Отзывы: