background image

3

2.  Remove the end cover from the electrical compartment by

undoing the retaining screws in the left hand panel.

3.

Take the metal stand and place in the desired location. Note
that 2 screw holes are provided for wall mounting. If wall
mounted the unit should be secured using 38mm (1.5”), No.
10 screws.

4.

Take the boiler, the heater and the gasket kit and assemble as
shown in 

figure 2

.

i)

Place the bolts through the metal flange in the manner
shown and then place the metal flange over the tapered
glass tube of the boiler.

ii) 

Take the plastic insert and bend it round the glass tube
and into the metal flange. Pull the flange and insert
towards the open end of the glass tube until the insert
presses on the glass.

iii)  Place the aluminium plate, followed by the silicone ring

over the heating element and locate approximately 40-
45mm from the end with the electric cable.

iv)  Insert the heating element through the tapered glass tube

and into the boiler – ensure that there is clearance at the
inside end and no danger of the end of the element
pressing against the end of the boiler.

v) 

Now secure the element using the 3 bolts provided. Do
not overtighten.

Location

The Distinction Water Still can be wall or bench mounted. Select a
convenient location which has access to the following services:-

Electricity Supply

Before connection please ensure that the line supply is suitable. The
Distinction Water Still is available in two models, model no. D4000
for supplies rated at 3kW, 220-240V 50/60 Hz single phase and
model no. D4000/EURO for supplies rated at 3kW, 200-230V,
50/60Hz, single phase.

Water Supply

A cold water supply capable of providing a minimum flow rate of
60 l/hr.

Drain

A waste water drain located below the level of the Still so that the
drain pipe can fall away straight without kinks or bends, to allow an
unimpeded flow.

Reservoir

A distillate collection reservoir should be located beneath the Still.

Assembly

1. 

Unpack the Still and identify the following components:-

Qty. Component

Catalogue 

No.

1 Support 

Stand 

D4000/S

1 Boiler 

D4000/B

1 Condenser 

D4000/M2

Heating element 

D240H for D4000

D220H for D4000/EURO

1 Hose 

kit 

W4000/HK

1 Gasket 

kit 

D4000/GK

1 Boiler 

thermostat 

tube 

WTT48

1 Condenser 

thermostat 

tube

A515414

Plastic connecting caps 

QC18/11

Silicone rubber rings 

QR18/7 and QR18/10

1 P.T.F.E 

washer 

QW18/7

Reservoir level control 

WLS

Figure 2  Heater and thermostat tube assembly

Heater

Aluminium

plate

Silicone

ring

Plastic

insert

Metal

flange

Bolts

Boiler

Screwcap

Tapered glass tube

Boiler

thermostat

tube

40-45mm

Содержание Distinction D4000

Страница 1: ...Distinction Assembly and operating instructions Instructions de montage et de fonctionnement D4000 D4000 EURO Water Still Distinction...

Страница 2: ...ith or be affected by other equipment with similar qualifications We cannot be sure that other equipment used in its vicinity will meet these standards and we cannot guarantee that interference will n...

Страница 3: ...n Water Still can be wall or bench mounted Select a convenient location which has access to the following services Electricity Supply Before connection please ensure that the line supply is suitable T...

Страница 4: ...struct access to the thermostat reset buttons see figure 1 5 Take the two metal straps and attach the springs to the rear of the cradles on the stand Place the assembled boiler and element onto the cr...

Страница 5: ...d use 1 5mm M4 eyelet crimps with a performance conforming to BS 4G178 part 1 1984 and part 2 1986 A suitable crimp tool should be used Operation 1 Turn on the cold water supply and adjust the flow to...

Страница 6: ...SCALING OPERATION REMOVE ANY ACID SPILLS IMMEDIATELY Turn on the water supply and fill the boiler to the normal operating level The water will flush the acid into the boiler The water supply should be...

Страница 7: ...Failed electricity supply Restore supply c Thermostat operated Identify and rectify the cause and reset thermostat Water level in boiler a Supply of cooling water is Adjust the supply to is too low h...

Страница 8: ...de fr quences radio Il ne doit donc pas interf rer avec d autres appareils r pondant aux m mes directives Ne pouvant tre s r que d autres appareils en fonctionment autour du distillateur soint confor...

Страница 9: ...2 i Ins rer les boulons au travers du collier m tallique et le placer autour du tube vas du bouilleur ii Enrouler le collier en fibre et l ins rer entre le tube vas et le collier m tallique Le pliage...

Страница 10: ...collier plastique fourni 10 Placer l extr mit libre du tuyau vers l vacuation en s assurant qu il ne pr sente aucun coude pouvant perturber l coulement de l eau 11 V rifier que le robinet du c ntroleu...

Страница 11: ...pince sertir est n cessaire Mise en route 1 Ouvrir le robinet d alimentation en eau froide et r gler le d bit 1 l mn environ L eau doit circuler travers le r frig rant Attendre que le niveau d eau dan...

Страница 12: ...TECTION DES GANTS ET DES MASQUES FACIAUX DEVRONT ETRE UTILISES LORS DES OPERATIONS DE NETTOYAGE NETTOYER IMMEDIATEMENT TOUTE ECLABOUSSURE D ACIDE Ouvrir l alimentation d eau pour remplir le bouilleur...

Страница 13: ...tion du distillat vers V rifier que le tuyau de sortie du vers l vacuation le r servoir se fait mal distillat ne pr sente aucun coude b Event obstru par la sonde du D gager l vent thermostat r frig ra...

Страница 14: ......

Страница 15: ...Notes...

Страница 16: ...f bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketi...

Отзывы: