2

Before use

If the equipment is not used in the manner described in this
manual the protection provided by the equipment may be
impaired.

N.B. The Distinction Water Still is classified as “Permanently
Connected Equipment” and should be connected to the
electricity supply by a qualified electrician in the manner
described in the electrical installation section of this manual.

The Distinction Water Still is designed to operate under the
following conditions:

❖  

For indoor use only

❖  

Ambient temperature range +5°C to +40°C

❖  

Altitude up to 2000m

❖  

Relative humidity not exceeding 80%

❖  

Mains supply voltage fluctuations not greater than 
±10% of nominal

❖  

Overvoltage category II IEC60364-4-443

❖  

Pollution degree 2 IEC664

Limitations of use

This product meets the applicable E.C. harmonised standards for
radio frequency interference and may be expected not to interfere
with, or be affected by other equipment with similar qualifications.
We cannot be sure that other equipment used in its vicinity will meet
these standards and we cannot guarantee that interference will not
occur in practice. Where there is a possibility that injury, damage or
loss might occur if equipment malfunctions due to radio frequency
interference, or for general advice before use, please contact the
Technical Service Department of Bibby Scientific Ltd.

Distinction Water Still 
D4000 & D4000/Euro

The Distinction Water Still complies with the 
following European Directives:

89/336/EEC

- E.M.C. DIRECTIVE

Amended by 92/31/EEC and 93/68/EEC

73/23/EEC

- L.V.D. Amended by 93/68/EEC

Figure 1 General assembly

Thermostat reset buttons

Boiler
thermostat
tube

On/Off switch

Clip

Reservoir
level
control

Heater

Gaskets

and flange

Distillate

outlet tube

Boiler

Stand

Stopcock

To drain

Constant
level control

Wall mounting
screw holes

Cooling water
inlet

Condenser
thermostat tube

Condenser
outlet

Straps

Condenser

Condenser
thermostat

Boiler

thermostat

Distinction

Содержание Distinction D4000

Страница 1: ...Distinction Assembly and operating instructions Instructions de montage et de fonctionnement D4000 D4000 EURO Water Still Distinction...

Страница 2: ...ith or be affected by other equipment with similar qualifications We cannot be sure that other equipment used in its vicinity will meet these standards and we cannot guarantee that interference will n...

Страница 3: ...n Water Still can be wall or bench mounted Select a convenient location which has access to the following services Electricity Supply Before connection please ensure that the line supply is suitable T...

Страница 4: ...struct access to the thermostat reset buttons see figure 1 5 Take the two metal straps and attach the springs to the rear of the cradles on the stand Place the assembled boiler and element onto the cr...

Страница 5: ...d use 1 5mm M4 eyelet crimps with a performance conforming to BS 4G178 part 1 1984 and part 2 1986 A suitable crimp tool should be used Operation 1 Turn on the cold water supply and adjust the flow to...

Страница 6: ...SCALING OPERATION REMOVE ANY ACID SPILLS IMMEDIATELY Turn on the water supply and fill the boiler to the normal operating level The water will flush the acid into the boiler The water supply should be...

Страница 7: ...Failed electricity supply Restore supply c Thermostat operated Identify and rectify the cause and reset thermostat Water level in boiler a Supply of cooling water is Adjust the supply to is too low h...

Страница 8: ...de fr quences radio Il ne doit donc pas interf rer avec d autres appareils r pondant aux m mes directives Ne pouvant tre s r que d autres appareils en fonctionment autour du distillateur soint confor...

Страница 9: ...2 i Ins rer les boulons au travers du collier m tallique et le placer autour du tube vas du bouilleur ii Enrouler le collier en fibre et l ins rer entre le tube vas et le collier m tallique Le pliage...

Страница 10: ...collier plastique fourni 10 Placer l extr mit libre du tuyau vers l vacuation en s assurant qu il ne pr sente aucun coude pouvant perturber l coulement de l eau 11 V rifier que le robinet du c ntroleu...

Страница 11: ...pince sertir est n cessaire Mise en route 1 Ouvrir le robinet d alimentation en eau froide et r gler le d bit 1 l mn environ L eau doit circuler travers le r frig rant Attendre que le niveau d eau dan...

Страница 12: ...TECTION DES GANTS ET DES MASQUES FACIAUX DEVRONT ETRE UTILISES LORS DES OPERATIONS DE NETTOYAGE NETTOYER IMMEDIATEMENT TOUTE ECLABOUSSURE D ACIDE Ouvrir l alimentation d eau pour remplir le bouilleur...

Страница 13: ...tion du distillat vers V rifier que le tuyau de sortie du vers l vacuation le r servoir se fait mal distillat ne pr sente aucun coude b Event obstru par la sonde du D gager l vent thermostat r frig ra...

Страница 14: ......

Страница 15: ...Notes...

Страница 16: ...f bibby scientific fr www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail marketi...

Отзывы: