background image

12

608.25_TL70_2019-05-22_01_IM2_BEU_EN

 

MedNet GmbH, Borkstrasse 10, 48163 Münster, Germany

Distributed by: Beurer GmbH  •  Söflinger Str. 218  •  89077 Ulm, Germany

 

Tel.: +49 (0)731 / 39 89 -144  •  Fax: +49 (0)731 / 39 89 - 255

 

www.beurer.com • Mail: [email protected]

 

Dong Guan Q&S Electronic Manufacturing Company Ltd.

 

Yin Shan District, Fu Gang Village, Xiang Mang West Road, Qing Xi Town, Dong Guan City, P.R. China

0598

EC|REP

Table 4: Recommended separation distances between portable and mobile RF communications 

equipment and the model TL 70

Recommended separation distances between 

portable and mobile RF communications equipment and the model TL 70

The model TL 70 is intended for use in electromagnetic environment in which radiated RF disturbances is 

controlled. The customer or the user of the model TL 70 can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) 

and the model TL 70 as recommended below, according to the maximum output power of the communica-

tions equipment.

Rated maximum output 

power of transmitter 

W

Separation distance according to frequency of transmitter 

m

150 kHz to 80 MHz

1.2   P

80 MHz to 800 MHz

1.2   P

800 MHz to 2,5 GHz

2.3   P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 

in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where 

P

 is 

the maximum output power rating of the transmitter in watts (W) accordable to the transmitter manufacturer.

NOTE 1

 At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies. 

NOTE 2

  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-

on and reflection from structures, objects and people.

Содержание TL 70

Страница 1: ...EN Daylight therapy lamp Instruction for use 2 ELECTROMAGNETIC COMPATIBILITY INFORMATION 9 TL 70 0598...

Страница 2: ...ss This hormone tells the body when it is time to sleep In the months when there is less sunlight therefore production of melatonin increases This makes it harder to get up in the morning since the bo...

Страница 3: ...ble to other users and observe the information they contain Safety notes Warning The light of the daylight therapy lamp is only intended for use on the human body Before use ensure that all packaging...

Страница 4: ...unit can be dangerous and may cause injury or improper results The main material of the case is made of PC ABS Be cautious of potential allergic reactions to these ma terials General notes Important...

Страница 5: ...t 6 Operation 1 Insert the mains part into the connector 2 Switch on the light Press the ON OFF button 3 LED display timer The daylight therapy lamp TL 70 indicates your current treatment time with th...

Страница 6: ...n the distance Distance Duration Up to 10 cm hour day Up to 25 cm 1 hour day Up to 40 cm 2 hours day Basically the closer you are to the source of light the less time you should use it 7 Enjoying ligh...

Страница 7: ...m Operating conditions 5 C to 35 C 15 90 relative humidity atmospheric pressure range 86 106 kpa Storage conditions 20 C to 50 C 15 90 relative humidity atmospheric pressure range 86 106 kpa Product c...

Страница 8: ...first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g where they can send the...

Страница 9: ...t Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The models TL 70 use RF energy only for its internal function Therefore its RF emissions are very low and are not likely...

Страница 10: ...cal commercial or hospital environment Surge IEC 61000 4 5 1 kV line to line 2 kV line to earth 2 kV line to earth Mains power quality should be that of a typical commercial or hospital environment Vo...

Страница 11: ...trengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equip...

Страница 12: ...ing a minimum distance between portable and mobile RF communications equipment transmitters and the model TL 70 as recommended below according to the maximum output power of the communica tions equipm...

Отзывы: