Table 3: Guidance & Declaration - electromagnetic immunity concerning Conducted RF & Radiated RF

Guidance and manufacturer’s declaration – electromagnetic immunity

The model TL 70 is intended for use in the electromagnetic environment specified below.  

The customer or the user should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Conducted RF lEC 

61000-4-6 

Radiated RF lEC 

61000-4-3 

3 V

rms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 V

3 V/m

Portable and mobile RF communications 

equipment should be used no closer to any 

part of the model TL 70, including cables, 

than the recommended separation distance 

calculated from the equation applicable to 

the frequency of the transmitter.

Recommended separation distance

3V

d=1.2   P   80MHz to 800MHz

d=2.3   P  800MHz to 2.5 GHz

where 

P

 is the maximum output power rating 

of the transmitter In watts (W) according to 

the transmitter manufacturer and 

d

 Is the 

recommended separation distance in meters 

(m). 

Field strengths from fixed RF transmitters, 

as determined by an electromagnetic site 

survey,

a

 should be less than the compliance 

level in each frequency range.

b

Interference may occur In the vicinity of 

equipment marked with the following sym-

bol: 

symbol:

   

NOTE 1

 At 80 MHz end 800 MHz. the higher frequency range applies.

NOTE 2

  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-

on and reflection from structures, objects and people.

a  

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoreti-

cally with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electroma-

gnetic site survey should be considered. If the measured field strength in the location in which the model TL 

70 is used exceeds the applicable RF compliance level above, the model TL 70 should be observed to verify 

normal operation. If abnormal performance is observed, additional measures may be necessary, such as 

reorienting or relocating the model TL 70.

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Содержание TL 70

Страница 1: ...EN Daylight therapy lamp Instruction for use 2 ELECTROMAGNETIC COMPATIBILITY INFORMATION 9 TL 70 0598...

Страница 2: ...ss This hormone tells the body when it is time to sleep In the months when there is less sunlight therefore production of melatonin increases This makes it harder to get up in the morning since the bo...

Страница 3: ...ble to other users and observe the information they contain Safety notes Warning The light of the daylight therapy lamp is only intended for use on the human body Before use ensure that all packaging...

Страница 4: ...unit can be dangerous and may cause injury or improper results The main material of the case is made of PC ABS Be cautious of potential allergic reactions to these ma terials General notes Important...

Страница 5: ...t 6 Operation 1 Insert the mains part into the connector 2 Switch on the light Press the ON OFF button 3 LED display timer The daylight therapy lamp TL 70 indicates your current treatment time with th...

Страница 6: ...n the distance Distance Duration Up to 10 cm hour day Up to 25 cm 1 hour day Up to 40 cm 2 hours day Basically the closer you are to the source of light the less time you should use it 7 Enjoying ligh...

Страница 7: ...m Operating conditions 5 C to 35 C 15 90 relative humidity atmospheric pressure range 86 106 kpa Storage conditions 20 C to 50 C 15 90 relative humidity atmospheric pressure range 86 106 kpa Product c...

Страница 8: ...first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g where they can send the...

Страница 9: ...t Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The models TL 70 use RF energy only for its internal function Therefore its RF emissions are very low and are not likely...

Страница 10: ...cal commercial or hospital environment Surge IEC 61000 4 5 1 kV line to line 2 kV line to earth 2 kV line to earth Mains power quality should be that of a typical commercial or hospital environment Vo...

Страница 11: ...trengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur In the vicinity of equip...

Страница 12: ...ing a minimum distance between portable and mobile RF communications equipment transmitters and the model TL 70 as recommended below according to the maximum output power of the communica tions equipm...

Отзывы: