1
BETACONNECT
TM
Instructions for use
Language: English Version: 2013-10
Страница 1: ...1 BETACONNECTTM Instructions for use Language English Version 2013 10...
Страница 2: ...mes 5 2 3 Charging the BETACONNECTTM 6 2 4 Care storage and disposal 6 2 5 Cleaning 7 Chapter 3 Injection Process 8 3 1 Preparing BETACONNECTTM for injection 8 3 2 Injection procedure 9 3 3 Removing t...
Страница 3: ...o extreme heat or fire Avoid storage in direct sunlight The BETACONNECT TM is not recommended for the blind or visually impaired without the assistance of a sighted individual trained to use it Keep t...
Страница 4: ...nded Use The re usable electronic autoinjector is intended to help the user by automating subcutaneous injection using a user filled disposable syringe provided in the Betaseron package The BETACONNEC...
Страница 5: ...e needle the syringe is pushed forward so that the needle penetrates the skin Phase II Injecting the medication the medication is injected subcutaneously injection speed depends on injection speed set...
Страница 6: ...nge of 700 hPa to 1 060 hPa 525 mmHg to 795 mmHg Your BETACONNECT TM can be operated at a temperature from 5 to 40 C 41 104 F at a humidity level of 15 to 93 relative humidity not condensing and an at...
Страница 7: ...lcohol wipe In case medication is spilled inside the BETACONNECT TM it must be removed using a damp cloth Other parts of the BETACONNECT TM do not normally need to be cleaned if needed lightly dampen...
Страница 8: ...or one injection but the BETACONNECT TM must be recharged right after A flashing orange bar in the light bar indicates the battery level is too low to perform an injection and the BETACONNECT TM will...
Страница 9: ...t the injection site Gently hold the BETACONNECT TM at a 90 angle against the skin to activate the safety release The injection button illuminates with a flashing blue light when the safety is release...
Страница 10: ...eedle the syringe and needle are withdrawn after the injection Upon injection completion the BETACONNECT TM will have two short beeps and a flashing blue light bar The BETACONNECT TM will automaticall...
Страница 11: ...BETACONNECT TM with your other hand while removing the syringe Important Be careful not to get injured by the needle when you remove the empty syringe Do not put the needle cap back on the syringe Al...
Страница 12: ...he slider below the 8mm setting 4 2 Adjusting the injection speed Step 1 Open the lid fully The setting allows slow medium and fast injections By default BETACONNECT TM is set to the medium injection...
Страница 13: ...d Step 2 Set the reminder function to the desired setting Adjustment Reminder function enabled Reminder function disabled Step 3 Reminder feature The reminder is an intermitting beep and a blue flashi...
Страница 14: ...cannot be opened during the injection process 1 Wait until the injection process is completed 2 The lid or the opening button does not function 2 Please contact your MS nurse healthcare professional f...
Страница 15: ...terrupted ensure the entire dose is injected Consult your MS nurse healthcare professional if the entire dose was not received 2 A malfunction was detected 2 Hold the on off button down to shut down B...
Страница 16: ...and contribute to the prudent and rational use of natural resources The BETACONNECT TM is a type BF device and provides protection against electrical shock and electrical current leakage Manufacturer...
Страница 17: ...e BETACONNECT TM outside such an environment Mobile phones should depending on their transmitting power be kept at a distance of at least 1 m 3 28 ft from the BETACONNECT TM while it is in use Radio t...
Страница 18: ...local representative or your MS nurse healthcare professional if you have any questions on the BETACONNECT TM if you do not need it anymore if you want to dispose of it or in case of malfunctions We...