MasterCAL
Users Manual
68
Simultaneous Measure/Source
Use the MEASURE/SOURCE mode to calibrate or
emulate a process instrument. Press
M
so that a
split screen display appears as shown below.
gj42s.eps
Table 7 shows the functions you can use
simultaneously when Loop Power is disabled. Table
8 shows the functions you can use simultaneously
when Loop Power is enabled.
You can use the
Step
or
Auto Step
features to adjust
the output in MEASURE/SOURCE mode, or you can
use the calibration routine provided when you press
the
As Found
softkey.
The following two softkeys displayed in
MEASURE/SOURCE mode are for use in calibrating
a process instrument.
•
As Found
, which lets you set up a calibration
routine to obtain and record as found data.
•
Auto Step
, which lets you set up the calibrator for
auto-stepping, as previously described.
Instructions for calibrating a process instrument
follow.