INSTANT FLEXIBLE DIGITAL THERMOMETER

OWNER S M

   ANUAL

Model:DMT-4318

Warning : 

 

Read instructions thoroughly before using digital thermometer.
Choking Hazard:  Thermometer cap and battery may be fatal if swallowed.  Do not allow children
to use this device without parental supervision.The Patient is an intended operator.
Do not use thermometer in ear.  Designed use is for oral, rectal, and armpit (axilla) readings only.
Do not place thermometer battery near extreme heat as it may explode.
Remove battery from the device when not in operation for a long time.
The  use of temperature readings for self-diagnosis is dangerous.  Consult your doctor for the
interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions.
Do not attempt measurements when the thermometer is wet as inaccurate readings may result.
Do not bite the thermometer. Doing so may lead to breakage and/or injury.
Do not attempt to disassemble or repair the thermometer. Doing so may result in inaccurate readings.
After each use, disinfect the thermometer especially in case the device is used by more than one person.
Do not force the thermometer into the rectum. Stop insertion and abort the measurement when pain is
present. Failure to do so may lead to injury.
Do not use thermometer orally after being used rectally.
For children who are two years old or younger, please do not use the devices orally. 
If the unit has been stored at temperatures over 41

℉~

104

(5

℃~

40

℃)

, leave it in 41

℉~

104

(5

℃~

40

℃)

 ambient temperature for about 15 minutes before using it. 

  

Figure 2

2.   Position thermometer in desired location (mouth, rectum, or armpit.)  
     a)   

Oral Use:

  Place thermometer under tongue as indicated by

“ √ ”

           position shown in Figure 2.  Close your mouth and breathe evenly 
           through the nose to prevent the measurement from being influenced 
           by inhaled/exhaled air. 
     b)   

Rectal Use:

  Lubricate silver probe tip with petroleum jelly for easy insertion.  Gently insert 

           sensor approximately 1cm (less than 1/2 ") into rectum. 
     c)   

Armpit Use: 

 Wipe armpit dry.  Place probe in armpit and keep arm pressed firmly at side.  

          From a medical viewpoint, this method will always provide inaccurate readings, and should not 
          be used if precise measurements are required. 
3.   The degree sign flashes throughout the testing process.  When flashing stops an alarm will beep for 
     approximately 10 seconds.  The measured reading will appear on the LCD simultaneously. The 
     minimum measurement time until the signaling tone (beep) must be maintained without exception. The 
     measurement continues even after the buzzer notification. So that in order to achieve better body 
     temperature measurement result, recommend to keep the probe in mouth and rectum about 2 minutes,
     or in armpit about 5 minutes regardless of the beep sound and at least 30 seconds measurement interval
     should be maintained.

4.   To prolong battery life, press the 

On/Off

  Button to turn unit off after testing is complete.  If no action 

      is taken, the unit will automatically shut off after around10 minutes.

*Note:  Normally 

Alarm b eeps m ore ra pidly w hen te mperature  

the buzzes are

 Bi

-

Bi

-

Bi

-

B

   i

-

;   

              reaches 100

 (37.8

) or higher ,and t

he  buzzes are

 Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi

              

CLEANING AND DISINFECTION

1)Immerse the thermometer probe in distilled water for at least 1 minute;
2)Using a clean, soft cloth to wipe the thermometer down to remove any residue;
3)Repeat step 1 and 2 for three times until no soil is seen with visual inspection after cleaning;
4)For thoroughly clean and disinfection, please use method A or B

  

Method A(High level disinfection): immerse the thermometer probe in 0.55% OPA(O-Phthaldehyde),

    such as CIDEX OPA, for at least 12 minutes under temperature at 20

℃;

  

Method B(Low level disinfection): Using a clean soft cloth dipped in 70

medical alcohol

,

wipe the probe

    3 times

,

at least one minute for each time

.

5)Repeat step 1 to 3 to remove OPA residuals;
Note1: Rectal use is not recommended for home use as OPA will not be readily available outside of a hospital.
If rectal measurement is necessary, we strongly recommend high level disinfection.
Note2: Please operate according to the manual of OPA for reference.
To prevent damage to the thermometer please note and observe the following:
-Do not use benzene, paint thinner, gasoline or other strong solvents to clean the thermometer.
-Do not attempt to disinfect the sensing probe (tip) of the thermometer by immersing in alcohol, OPA
 or in hot water (water over 122

(50

) for long time

.

-Do not use ultrasonic washing to clean the thermometer.

Figure 4

Figure 5

Figure 3

 

 BATTERY REPLACEMENT

1.   Replace battery when

appears in the lower right corner of LCD display.

      

2.   Pull battery cover off as shown in Figure 3.
3.   Gently pull out plastic circuit board with battery chamber approximately 1 cm (slightly less than1/2

 

".)

     (See Figure 4)
4.   Use pointed object such as a pen to remove old battery.  Discard battery lawfully.  Replace with new 
     1.5V DC button type LR41 or SR41,UCC392, or equivalent.  Be sure battery is installed with 

+

     

polarity facing up.  (See Figure 5)

5.   Slide battery chamber back into place and attach cover.

TROUBLESHOOTING

Problem

Solution

Error message              

The system is not 
functioning properly.

Unload the battery, wait for 1 minute and repower
it. If the message reappears, contact the retailer 
for service.

Temperature taken is
lower than 90.0°F(32.0°C)

Turn off, wait one minute and take a new temperature
 via close contact and sufficient rest.

Temperature taken is 
higher than 109.9°F(42.9°C)

Dead battery: Battery icon is 
flashing, can’t be measurable. Replace the battery.

Turn off, wait one minute and take a new temperature
 via close contact and sufficient rest.

LIMITED WARRANTY

The thermometer  is  guaranteed for one year  from the date of  purchase. If  the  thermometer does  not 
function  properly due to defective components or poor workmanship, we will repair or replace it  free of
charge.  All components are covered by this warranty excluding the battery. The warranty does not  cover 
damages  to your thermometer due to improper handling.  

To obtain warranty service, an original or copy 

of the sales receipt from the original retailer is required.

Disposal of this product and used batteries should be carried out 
in accordance with the national regulations for the disposal of 
electronic products.

FCC INFORMATION

Caution: Changes or modifications to this unit not expressly approved by the party responsible for 
compliance could void the user authority to operate the equipment.  
*Note: 
This equipment has been tested and found to comply with the limits for a Class B digital device, 
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against 
harmful interference in a residential installation. This equipment generates, uses, and can radiate radio 
frequency energy. If this equipment does cause harmful interference to radio or television reception, 
which can be determined by turning the equipment off and on, the user is encouraged to try and correct 
the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna. 
- Increase the distance between the equipment and the receiver. 
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected. 
- Consult the dealer or an experienced radio/TV technician for help. 

Made in China

Document No. :J

DMT-2104-002

Article No.: 001
Version: A
Date of Issue: 2020.01

CALIBRATION

The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to 
the use instruction, periodic readjustment is not required. However, we recommend checking calibration 
every two years or whenever clinical accuracy of the thermometer is in question. Turn on the thermometer 
and insert into the water bath and then check the laboratory accuracy. Please send the complete device to the
dealers or manufacturer. ASTM laboratory accuracy requirements in the display range of 98.6 to 102.2 °F
(37.0 to 39.0 °C) for electronic thermometers is ±0.2°F(±0.1°C).
The above recommendations do not supersede the legal requirements. The user must always comply with 
legal requirements for the control of the measurement, functionality, and accuracy of the device which 
are required by the scope of relevant laws, directives or ordinances where the device is used.

JOYTECH HEALTHCARE  CO.LTD.
No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city, 311100 Zhejiang,China

Battery Cap

On/Off

  Button

LCD

Probe

Figure 1 

DIRECTIONS

1.   Press the

 

On/Off  

Button next to LCD display.  A tone will sound as the screen shows              , 

     followed by last recorded temperature. After showing the self-test temperature, the thermometer is 
     now in the testing mode.

Measure Range:
Accuracy:

Operating mode:                        Direct Mode
Display: 
Memory:
Battery: 
Battery life: 

Dimension: 
Weight: 

90.0

-109.9

(32.0

℃-

42.9

 ) (

  / 

c h

  osen b y m anufacturer )  

±

0.2

(

±

0.1

) during 95.9

~107.6

(35.5

~42.0

) at

64.4

~82.4

(18

~28

 ) ambient operating range

±

0.4

(

±

0.2

) for other measuring and ambient operating range

Liquid crystal display, 3 1/2 digits 
For storing the last measured value
One 1.5 V DC. button battery (size LR41or SR41, UCC 392) 

13.9cm x 2.2cm x 1.3cm (L x W x H) 

Approx. 11 grams including battery 

Type:

Digital Thermometer (Not Predictive)

PRECAUTION

*

The performance of the device may be degraded should one or more of the following occur:

- Operation outside the manufacturer’s s tated te mperature and humidity range.
- Storage outside the manufacturer’s s tated te mperature and humidity range.
- Mechanical shock (for example, drop test) or degraded sensor.
- Patient temperature is below ambient temperature.
*Portable and mobile RF communications can affect the device. The device needs special pre-cautions
   regarding EMC according to the EMC information provided in the accompany documents.
*Do not use the devices in the MR environment.

CONTENTS

1 Thermometer,  1 Owner s Manual,  1 Storage Case

SPECIFICATIONS

SYMBOL EXPLANATION

 Direct Current 

 Type BF  Applied Part

Batch Code

Manufacturer

PLEASE  READ  CAREFULLY  BEFORE  USING

This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. To better understand its functions and to provide years of dependable
results, please read all instructions first.
This appliance conforms to the following standards:
ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of
Patient Temperature,
ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular requirements for basic safety
and essential performance of clinical thermometers for body temperature measurement,
IEC 60601-1-11 Medical electrical equipment —Part 1-11: General requirements for basic safety and
essential performance –Collateral Standard: Requirements for medical electrical equipment and 
medical electrical systems used in the home healthcare environment and complies with the 
requirements of IEC 60601-1-2(EMC), AAMI/ANSI ES60601-1(Safety) standards. And the 
manufacturer is ISO 13485 certified.

 

Consult Accompanying Documents

Storage and Transportation 
Temperature Limit:-4

℉ ~

131

 (-20

~55

℃)

-20

55

Approx. 200hours of continuous operation or 1 year with 3 
measurements per day

Ambient operating range:

Storage and transportation
condition:  

Temperature: 41

℉~

104

(5

℃~

40

)

Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa

Temperature:-4

℉~

131

 (-20

~55

)

Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa

Classification:

Type BF

Ingress Protection Rating: 

IP 27

Expected service life:                Three years

/

 SWITCHABLE 

Temperature readings are available in the Fahrenheit or Celsius scale (

/

; located in the upper right 

corner of LCD.)  With the unit off, press and hold the

 On/Off  

Button for approximately 2 seconds to

change the current setting.

The digital thermometers are intended to measure the human body temperature in regular
mode orally, rectally or under the arm. And the devices are reusable for clinical or home use on people of
all ages, including children under 8 years old with adult supervision. 

Indications For Use