29
WWW.BERTIN-INSTRUMENTS.FR
USER MANUAL
Use a wipe dampened with a surfactant-water solution to clean the external
parts of the Coriolis Compact unit
If the power supply cable has been in contact with the contaminated
environment, clean it in the same way
Wipe down the assembly using a soft cloth
Decontamination
If samples have been obtained in a zone that is contaminated by biological
agents, a decontamination step is required after cleaning.
The decontamination procedure detailed below is that implemented by Bertin
during testing. It is recommended that you adapt the decontamination procedure
according to the risk of infection.
Spray the external parts of the Coriolis Compact with 12° chlorometric bleach
Leave to incubate for 30 minutes
Rinse carefully with fresh water
Dry using a soft cloth
Note 1: decontamination must only be performed on equipment items that have
already undergone cleaning.
Note 2: decontamination must not be performed multiple times in succession to
avoid damaging the unit.