7 Main Menu Device Setup
LB 470 Level
140
56925BA2 Rev. 03, 11/2017
Analog Output: Calibrate
Device settings | Setup | Output | Analog Output (AO) | AO Calibration
If there are any discrepancies between the target value and the actual value of the
current signal, then the current output may be calibrated again.
NOTICE
For calibration of the current output, an ammeter (not included in the scope of
delivery) is required, which is connected to the current output.
NOTICE
BERTHOLD TECHNOLOGIES recommends calibrating the current outputs when-
ever a module has been installed/replaced or if a software update has been
carried out.
1
2
3
Execution with connection board
Execution with terminal block
Button
<Calibrate>
4
5
Input field 4 mA Calibration point
Input field 20 mA Calibration point
Fig. 123 Analog Output (Calibration)