5.1.4
Protective Earth and Equipotential Bonding
The protective earth conductor has to be connected to the terminals marked
with "PE".
The housing must be connected to local equipotential bonding.
5.1.5
EIA-485 (RS-485) Network
For integration of EVU units into an EIA-485 (RS-485) network, all participants must
be connected one after the other in the configuration Master-Master. Star connec-
tion is not permitted.
The first and last station (physical, independent of the master's position) on the
network needs a terminating resistor of 120 Ω.