Chapter 4 Getting Started
56
4.4
Calibration in Standard Mode
Empty calibration has to be performed in order to calibrate the
measurement.
Prerequisites (see also page
Container must be empty or at least below the monitoring
level.
The gas density existing under actual operating conditions is
available in the container. If this is not possible, select the pro-
fessional mode and compensate the gas density value with
code 35. The special features of the professional mode have to
be taken into account.
A possible existing heating or cooling jacket is filled with the
medium to be used later.
In agitator containers, the agitator must in operation, if it may
have any influence on the limit value.
The shielding with the source has been installed. The radiation
exit channel is open.
The EVU is not in the edit mode.
Calibration:
Push Cal button for 3 seconds.
The seconds are counted down in the top row.
The bottom row shows the count rate.
In code 20 you can read off the measured zero count rate.
The measurement is now calibrated and already sup-
plies live measured values.
Note!
A system check has to be carried out to ensure that the measure-
ment operates correctly.
To do this, fill the container above fill level limit, or alternatively,
close the shielding when the container is empty.
59s
24
Содержание LB4710-050
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 32: ......
Страница 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Страница 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Страница 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Страница 44: ......
Страница 58: ......
Страница 98: ......
Страница 104: ......
Страница 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Страница 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Страница 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Страница 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Страница 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Страница 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Страница 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Страница 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Страница 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Страница 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Страница 120: ......
Страница 141: ...Mini Switch LB 471 141 Notes...
Страница 142: ...Mini Switch LB 471 142 Notes...
Страница 145: ......
Страница 146: ...Limit Switch Mini Switch LB 471 Software Manual User s Guide detect and identify Process Control...
Страница 147: ......
Страница 149: ......
Страница 151: ......
Страница 161: ......
Страница 195: ......
Страница 215: ......
Страница 219: ......
Страница 223: ...Mini Switch LB 471 78 Notes...
Страница 224: ...Mini Switch LB 471 79 Notes...
Страница 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...