Note!
Leave the time constant on automatic unless there are good rea-
sons for entering a fixed value.
If you enter a fixed value as time constant, you should keep in
mind that a fixed value does not adjust to the source decay.
Therefore, enter a value that ensures a safe switching function
even after 10 years.
Rule of thumb: automatically calculated time constant x 10.
If the container is under high gas pressure or if significant wall
caking is to be expected, then it is absolutely essential that you
enter a fixed value. The value should be so great that possible gas
density variations or wall caking cannot trigger the alarm.
Rule of thumb: automatically calculated time constant x 30.
Live count rate (cps)
Display of non-average count rate.
(Only in the professional mode)
The displayed count rate comes directly from the detector and is
not averaged by the time constant.
Maximum time constant (s)
Upper limit of time constant in code 12
(Only in the professional mode)
Input
Maximum time constant
0
Disabled
> Code 12
Enabled
If the time constant in code 12 is set to automatic, then the time
constant in code 12 is automatically increased or adapted to the
source decay, so that no false alarm can be triggered by statistical
fluctuations. To avoid that the time constant may become too big
over the years, you can here enter a value that triggers an alarm
as soon as the time constant in code 12 exceeds this value. The
display shows the error message 39.01 "Replace source". Code 55
shows the year in which the message is likely to be output.
Please keep in mind: If a fixed value has been entered in code 55,
then the error message "Replace source" is displayed as soon as
this date is exceeded. Code 14 is then disabled.
C o d e 1 3
C o d e 1 4
Содержание LB4710-050
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 32: ......
Страница 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Страница 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Страница 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Страница 44: ......
Страница 58: ......
Страница 98: ......
Страница 104: ......
Страница 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Страница 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Страница 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Страница 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Страница 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Страница 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Страница 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Страница 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Страница 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Страница 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Страница 120: ......
Страница 141: ...Mini Switch LB 471 141 Notes...
Страница 142: ...Mini Switch LB 471 142 Notes...
Страница 145: ......
Страница 146: ...Limit Switch Mini Switch LB 471 Software Manual User s Guide detect and identify Process Control...
Страница 147: ......
Страница 149: ......
Страница 151: ......
Страница 161: ......
Страница 195: ......
Страница 215: ......
Страница 219: ......
Страница 223: ...Mini Switch LB 471 78 Notes...
Страница 224: ...Mini Switch LB 471 79 Notes...
Страница 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...