English
Audio ALARM
(optional factory setting)
In case of an upper or lower temperature excursion, 3
audible ALARM signals will be triggered immediately
(500ms ON/500ms OFF), thereafter as following:
During 10 minutes
→
Every minute 1 ALARM -signal
During 50 minutes
→
Every 10 minutes 1 ALARM -signal
During 11 hours
→
Every hour 1 ALARM -signal
Acknowledge audible ALARM: Press the
READ
button,
if the temperature is in the allowed temperature range.
13.3) Option 3:
Read out data from the
files generated by the Fridge-tag 3
ext. sensor with a computer
Plug the
Fridge-tag
®
3
into any computer via USB
Interface. Make sure the device is plugged in properly.
Wait sufficient time for the device to generate the ASCII
and PDF files (depending on the programming, this
process may take a couple of minutes). You can see
that the device is working from the continuously appea-
ring arrows in the upper display area.
Temperature / Min. - Max
parameter 2
parameter 1