Chapter 6 Instructions for use: Ikus
What?
When?
Meaning
Pump stands still - no
further pump function;
acoustic alarm is still
audible
Batteries are complete
emptied or there is a
serious fault in the
batteries
Immediately connect the
Ikus to the mains. Until
that supply the patient
with the manual pump.
Acoustic alarm is
audible; the displays are
off and it is not possible
to perform any action on
the laptop
Batteries are defective
or mains failure
Immediately connect the
Ikus to the mains. Until
that supply the patient
with the manual pump.
Tab. 6-4
Displays and messages during battery operation mode
Switching over to mains operation
INSTRUCTION
1.
Connect the Ikus to the mains.
2.
The indicator
Mains operation
lights up. The indicator
Battery operation
in the
handle goes out. The battery is now being charged.
6.5
Changing over from univentricular to biventricular
operation
CAUTION
After changing over to biventricular operation the device is operating in
separate mode. All parameters are reset to the default parameters (see
Tab. 15-3, page 203). The patient-customized parameters have to be
adjusted again.
INSTRUCTION
1.
The pump which is to be connected must be already de-aired and connected
to the cannulae.
2.
In the monitor program, select the option
Drive pause
(see Fig. 6-17,
page 77) and press
<Enter>
to confirm. Confirm the selection in the dialog
window by pressing
<X>
or
<1>
. The active pump will stop. The view
Select
operating mode
appears.
3.
Open the driving tube connector identified by a blue marking.
4.
When changing from univentricular support to BVAD: ensure the LVAD pump
is attached to the red connector and RVAD pump is attached to the blue con
nector. To do so, take hold of the plug's release sleeve and pull the plug out of
the socket.
5.
Plug the driving tube of the de-aired new pump into the free driving tube con
nector. The sound of the plug snapping into place is clearly audible. Check
that the plug is securely connected. To do so, grip the release sleeve above
the grooved section and pull on it. Do not pull from the release sleeve, and
never from the tube!
6.
Select
Biventricular (BVAD)
and confirm with
<Enter>
. All following steps are
the same as required for the start-up procedure. See chapter 7: Commission
ing the Ikus and setting parameters, page 93.
1000721x09 Revision 8
82
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...