Chapter 6 Instructions for use: Ikus
6.2.7 Adjusting the parameter values
NOTICE
If the adjusted parameter was not confirmed with
<Enter>
, the
parameter table will display the changed parameter until the automatic
log off, but the Ikus will continue operating with the former value.
The changed parameter display can be Ed with
<Esc>
. The original
value appears again in the parameter table.
To confirm the change, press
<Enter>
; the new value is displayed in
the parameter table.
INSTRUCTION
1.
Use the
<
←
>
/
<
→
>
keys to move the cursor to the desired field in the parame
ter table. The selected field is given a colored background.
2.
Use the
<
↓
>
,
<
↑
>
/
<Bild-
↓
>
,
<Bild-
↑
>
keys to adjust the value, then press
<Enter>
to confirm. The system will now operate using the new value.
3.
Visually check that the blood pump(s) is (are) filling and ejecting completely
over a period of several pump cycles!
Fig. 6-16
Parameter table
Parameter
Range possible
<
↓
>/<
↑
>
changes value
by
<Bild-
↓
>/<Bild-
↑
>
changes value by
Systolic pressure
[mmHg]
60 to 350
2.5
25
Diastolic
pressure [mmHg]
0 to -100
2.5
25
Rate [bpm]
30 to 150
1
10
Relative systolic
duration [%]
20 to 70
1
10
Tab. 6-3
Parameter adjustment
1000721x09 Revision 8
76
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...