Chapter 6 Instructions for use: Ikus
6.2.2 Shutting down the monitor program
WARNING
Do not shut down the monitor program unless this is absolutely
necessary (e. g. if new user profiles have to be set up). Restart the
monitor program as soon as possible.
INSTRUCTION
1.
Press
<F10>
and confirm by pressing
<X>
or
<1>
. The start menu is displayed
on the laptop. The monitor program has now been shut down.
When the monitor program has been shut down:
•
the system continues running with the currently set parameters
•
no data on current events are recorded, the LOG files will be incomplete
•
an acoustic signal and the indicator light on the handle control panel will alert the
user of new messages. However there is no way to display the message that has
been detected while the monitor program is shut down.
6.2.3 Logging in and out of the monitor program
IMPORTANT: This automatic forced logout does not occur after the first login during
the very first commissioning of the Ikus.
IMPORTANT: The Ikus performs a self-test of the alarm circuit for each login
procedure. If the self-test is completed successfully, the message
Acoustic alarm:
OK
appears a few seconds later. IMPORTANT: The self-test does not occur if, at the
time of login, an alarm is pending that was muted with the button
Mute alarm
(see
section 5.2.2: Display and operating panel, page 61).
IMPORTANT: If a problem arises during the self-test of the alarm circuit, the acoustic
signal continues for test purposes and cannot be muted during this phase.
Acknowledgment is only possible when a message appears (
Alarm circuit test failed
- buzzer remains off!
or
Acoustic alarm is not properly recognized
).
To select options in the monitor program, to change parameters or scroll through the
message window a user has to be logged into the monitor program. If no entries are
made for several minutes, the monitor program will log the user out automatically. The
Ikus continues to operate using the current parameter settings. The standard view is
displayed.
6.2.4 Logging in
Fig. 6-10
User ID
Fig. 6-11
Password
1000721x09 Revision 8
72
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...