Chapter 5 Description: Ikus
5.5.2 Redundant design of pneumatic systems in biventricular operation
Each blood pump is powered by 1 pneumatic systems. The 3
rd
system is redundant
and serves as a backup. If one of the active systems fails, the 3
rd
(backup) system
drives the corresponding pump. The performance of the Ikus remains the same. If 2
systems should fail, the remaining pneumatic system will drive both pumps. In this
case, the Ikus is running in emergency operating mode.
Normal operation
Failure of 1 system
Failure of 2 systems
System 1 left pump
System 1 failure
System 1 failure
System 2 right pump
System 2 right pump
System 2 failure
System 3 backup
System 3 left pump
System 3 left and right
pump
Message
Backup opera
tion left. Contact cos-
Message
Emergency
tumer service.
is
operation System 3.
displayed.
Contact service now!
is
displayed.
Fig. 5-7
Pneumatic system redundancy in biventricular operation
Emergency operating mode
If 2 pneumatic systems fail while in biventricular operation, the remaining system will
drive both blood pumps. In this case, the
Ikus
will be operated in synchronous mode
with a 250 mmHg systolic pressure, -100 mmHg diastolic pressure, 70 bpm and a
relative systolic duration period of 40 %.
5.5.3 Control computer with redundancy design
The control computers are also designed to provide backup redundancy. The Ikus
system has 2 independently operating control computers. The main computer controls
the pneumatic systems and transmits important function-specific data to the laptop
computer. The backup computer continuously compares its calculation results with
those of the main computer. If it detects a difference, an error message is generated.
For safety reasons, the control computers are located inside the
Ikus
casing and
operate fully independently of the laptop.
Emergency pulsing
The final safety system of the drive’s electronic system is the emergency pulse circuit
board. Should both control computers fail, this hardware takes over control of the
system. In order to preclude all possible sources of faults, the emergency pulse circuit
1000721x09 Revision 8
64
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...