Chapter 5 Description: Ikus
5.4
Laptop computer with monitor program
NOTICE
If the laptop is switched off while the Ikus is in operation, the driving unit
will continue to operate using the current parameter settings.
The laptop computer is integrated into the Ikus casing. The laptop, with its
permanently installed monitor program, is used to start, set up and monitor the
system.
The monitor program, when running, provides the user with a continuous stream of
data on the system’s condition, as well as information on events and function faults.
At the same time, relevant data is stored on the hard disk for subsequent evaluation.
A detailed description of the monitor program is given in chapter 6: Instructions for
use: Ikus, page 67.
Fig. 5-5
Laptop
Fig. 5-6
Monitoring program
5.5
Safety
5.5.1
Redundant design of pneumatic systems in univentricular
operation
The blood pump is powered by 1 pneumatic system. The other 2 systems are
redundant and serve as backups. If the active system fails, one backup system drives
the pump. There is still an additional, redundant system available.
1000721x09 Revision 8
63
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...