Chapter 15 Troubleshooting and correcting faults
15.7 Reading out the LOG files
This is necessary if it is not possible to clearly identify function faults even after
consultation with the service department.
WARNING
If the user exits the monitor program, it is not possible to identify any
incoming messages. For this reason, always start the monitor program
again immediately after saving the data.
Use the USB stick provided with the device to save the data. Do not
connect any other USB device to the laptop (e.g. wireless technology).
Make sure that you always have the USB stick inserted and that there
is sufficient capacity on the stick. Otherwise the LOG files might get lost
as they are deleted from the hard disk as soon as they have been
transferred onto a USB stick.
INSTRUCTION
1.
Press
<F10>
to exit the monitor program and confirm the intention in the dia
log window by pressing the
<X>
key or the
<1>
key. The start menu is dis
played. The Ikus will continue to operate using the current parameter settings.
2.
Switch off the laptop. Insert the USB stick into the port (never do this while the
laptop is switched on:). Switch on the laptop again.
3.
Select the
4. Save data
option in the start menu. The LOG files are saved onto
the USB stick. After completion the start menu appears.
4.
Switch off the laptop and remove the USB stick (never do this while the laptop
is still switched on!)
5.
Switch the laptop on again. To return to the monitor program select the 1. Start
program option in the start menu.
6.
Enter user ID and password, confirming the password with
<Enter>
.
7.
Send the LOG files by e-mail to [email protected].
Fig. 15-5
Laptop CF30 with inserted USB stick
1000721x09 Revision 8
207
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...