Chapter 9 Implantation - surgical procedure
9.8
Shortening the cannulae if necessary
WARNING
If an EXCOR cannula extension set / connecting set is required for
implantation and the length of the tube part needs to be reduced, the
tube part should be cut but only to achieve the following minimum
lengths:
Article
Diameter /
Diameter reduction
Minimum length
Cannula extension set
A06-006
6 mm
55 mm
A09-009
9 mm
60 mm
A12-012
12 mm
75 mm
Connecting set
A06-009
9 to 6 mm
60 mm
A09-012
12 to 9 mm
75 mm
Tab. 9-2
Cannula extension set / connecting set: minimum length of tube section
INSTRUCTION
1.
Cut the cannulae to the required length. Make the cut perpendicular to the
cannula axis and ensure that the cut is straight.
2.
Make sure that the lengths of the 2 cannulae leading to the same pump match.
It must be possible to connect the cannulae to the pump without having to
exert any tension.
9.9
Connecting the blood pumps to the cannulae
WARNING
Ensure that cannulae, blood pump(s) and driving tubes are not subject
to external forces and are free of kinks or sharp bends.
When connecting the blood pump(s), pay attention to the direction of
the arrows on the inflow and outflow stubs. These show the direction of
the blood flow.
Type of support
Anastomosis of inflow
cannula to
Points upwards...
Univentricular
LVAD
apex
blood chamber
LVAD
atrium
air chamber
Tab. 9-3
Anastomosis and direction of the blood chambers
1000721x09 Revision 8
128
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...