1015002x00
103
Configuring and Operating the Driving Unit
en
8.6.3
BVAD:
2.
Every start of a blood pump begins with a step. The first step is
always a diastole. If the membrane of the blood pump is in the
diastolic end position, membrane movement cannot be
observed.
Each time the
S
TEP
button is pressed, another step will be per-
formed. The blood pump empties / fills the blood chamber.
Pressing the button twice will yield a complete pump cycle con-
sisting of a diastole and a systole.
3.
Ensure complete de-airing in single step mode. Check connec-
tion tightness. Verify anastomosis as well as the connections
between the components (cannulae to blood pumps; connec-
tor sets and extension sets if applicable).
4.
After conducting one step, the blood pump can be started with
A
UTO
. The pump cycles are conducted in accordance with the
set parameters.
To stop the blood pumps, the
S
TART
/ S
TOP
button must be
pressed once again. Next, press the
S
TOP
button. The blood
pump can be restarted with
S
TEP
or
A
UTO
.
5.
The start of the pump is thus concluded. The parameters must
be adjusted subsequently. Make sure that the blood pump is
filling and emptying completely.
6.
Continue with chapter 8.6.4: Setting Rate and Synchroniza-
tion, page 104.
1
1.
Press the
S
TART
/ S
TOP
button.
2.
Every start of a blood pump begins with a step. The first step is
always a diastole. If the membrane of the blood pump is in the
diastolic end position, membrane movement cannot be
observed.
The left blood pump must be started before the right blood
pump.
Each time the
S
TEP
button is pressed, another step will be per-
formed. The blood pump empties / fills the blood chamber.
Pressing the button twice will yield a complete pump cycle con-
sisting of a diastole and a systole.
3.
Ensure complete de-airing in single step mode. Check that
connections are air-tight. Verify anastomosis as well as the
connections between the components (cannulae to blood
pumps; connector sets and extension sets if applicable).
4.
After each blood pump has performed at least one step, the left
blood pump or both blood pumps can be started using
A
UTO
.
The pump cycles are conducted in accordance with the set
parameters.
1
Содержание EXCOR Adult
Страница 2: ...2 1015002x00 en ...
Страница 12: ...12 1015002x00 Table of Contents en ...
Страница 14: ...14 1015002x00 Contact en ...
Страница 66: ...66 1015002x00 Driving Unit en ...
Страница 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Страница 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Страница 126: ...126 1015002x00 Replacing the Driving System en ...
Страница 128: ...128 1015002x00 Weaning and Explantation en ...
Страница 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Страница 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Страница 154: ...154 1015002x00 Manual Pump en ...
Страница 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Страница 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Страница 194: ...194 1015002x00 Technical Specifications en ...
Страница 198: ...198 1015002x00 Overview Parameters en ...
Страница 206: ...206 1015002x00 EMC en ...
Страница 208: ...208 1015002x00 Patient ID Labels en ...
Страница 210: ...Abbrevations 210 1015002x00 en ...
Страница 212: ...212 1015002x00 en ...