NOTICE
Reinforcement plate, threaded rods, washers, and hexagonal nuts are
not included in the scope of delivery and must be supplied by the custom‐
er. The threaded rods must be of a screw strength class of at least 8.8,
pursuant to DIN ISO 898 P1. The reinforcement plate can be purchased
from BERCHTOLD (Dimensional drawings of ceiling anchor plates and
spacer).
5.5.2 Anchor plate and spacer block with heavy-duty anchors
Install the anchor plate on the reinforced concrete ceiling, preferably with a re-
inforcement plate. If this is not possible, heavy-duty anchors may be used for in-
stallation on concrete ceilings under specified conditions to be tested on a
case-by-case basis. The following designs are valid for the Federal Republic of
Germany. In other countries the relevant national regulations must be observed.
Basic information
Examination light CHROMOPHARE F 300
Installation and initial start-up
32