Monitoring Pulse Oxygen Saturation (SpO2) (Optional)
Safety
TMS60 Telemetry Monitoring System Operator’s Manual
9 - 3
■
Presence of certain dyes, such as methylene and indigo carmine
■
Inappropriate positioning of the SpO
2
sensor, or use of incorrect SpO
2
sensor
■
Drop of arterial blood flow to immeaurable level caused by shock, anemia, low
temperature or vasoconstrictor.
9.3
Safety
WARNING
•
Only use SpO
2
sensors specified in this manual. Follow the SpO
2
sensor’s
instructions for use and adhere to all warnings and cautions.
•
The operator is responsible for checking the compatibility of the pulse
oximetry monitor, sensor, and patient cable prior to use. Incompatible
components can result in degraded performance and/or device mal-
function.
•
When a trend toward patient deoxygenation is indicated, blood samples
should be analyzed by a laboratory co-oximeter to completely under-
stand the patient’s condition.
•
Do not use SpO
2
sensors during magnetic resonance imaging (MRI).
Induced current could potentially cause burns. The sensor may affect
the MRI image, and the MRI unit may affect the accuracy of the oximetry
measurements.
•
Prolonged and continuous monitoring may increase the temperature of
the sensor and cause the patient discomfort. It is especially important to
check the sensor placement, and ensure proper attachment on patients
suffering from poor perfusion or skin sensitivity. Check the sensor loca-
tion every two to three hours and move to another location if the skin
deteriorates. More frequent examinations may be required for different
patients.
•
Check accessory compatibility prior to use. Incompatible accessories
may degrade the device’s performance and may cause harm to the
patient.
•
Do not use disposable SpO
2
sensor for multiple times. Otherwise, inac-
curate SpO
2
measurement may occur.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...