Introduction
Alarms
6 - 2
TMS60 Operator’s Manual
6.1
Introduction
Alarms, triggered by technical problems, are visually and audibly indicated to the user
when the display is on.
The TD60 provides a subset of the technical alarms, this chapter describes the technical
alarms presented at the TD60 only. For details regarding full list of the technical alarms
as well as the physiological alarms displayed at the Central Monitoring System (CMS),
refer to
"Physiological Alarms" on page 12 - 2
.
6.2
Alarm Categories
The TD60’s alarm system only supports technical alarms and does not provide visual or
audible physiological alarms at the Device. These physiological alarms can only be
viewed at the CMS. See
BeneVision Central Monitoring System Operator’s Manual
for
details on physiological alarms.
Technical alarms are triggered by system status, patient status, a device malfunction or a
patient data distortion due to improper operation or mechanical problems. Technical
alarms are available both at the TD60 and the CMS.
In addition to the technical alarms, the TD60 also displays informational message to
inform the user of patient/system status. The TD60 will display informational messages
in the message area at the top of the display.
6.3
Alarm Levels
The alarms can be classified into three severity levels: high level, medium level and low
level.
WARNING
•
The reception failure of alarm signals may occur in the distributed alarm
system.
Alarm Levels
Technical alarms
High level
Indicates a severe device malfunction or an improper operation, which could
make it possible that the monitor cannot detect critical patient status and
thus threaten the patient’s life, such as low battery.
Medium level
Indicates a device malfunction or an improper operation, which may not
threaten the patient’s life but may compromise the monitoring of vital physi-
ological parameters.
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...