Storing the Batteries
Battery
13 - 6
TMS60 Operator’s Manual
13.7
Storing the Batteries
13.7.1
Storing Rechargeable Lithium-ion Battery
When storing batteries, make sure that the battery terminals do not come into contact
with metallic objects.
If you need to store the batteries for an extended period of time, place the batteries in a
cool, dry place (ideally at 15°C or 60°F) with a partial charge of about 50% capacity (two
LEDs illuminated). Storing batteries in a cool place can slow the aging process.
Stored batteries should be charged to about 50% of their capacity every six months. The
battery should be fully charged prior to first use.
NOTE
•
Remove the lithium-ion battery from the device if the device is not used
for a prolonged time (for example, several weeks), and keep the device
in clean place to avoid the dust or liquid entering the battery compart-
ment.
•
Storing batteries at high temperatures for an extended period of time
will significantly shorten their life expectancy.
•
Do not store the batteries in an environment above 60°C (122°F) or lower
than -20°C (4°F).
Содержание TMS60
Страница 1: ...BeneVision TMS60 Telemetry Monitoring System Operator s Manual ...
Страница 2: ......
Страница 14: ...8 TMS60 Operator s Manual This page intentionally left blank ...
Страница 15: ...TMS60 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 7 ...
Страница 22: ...Equipment Symbols Safety 1 8 TMS60 Operator s Manual NOTE Some symbols may not appear on your equipment ...
Страница 50: ...Using the Pouch Getting Started 3 18 TMS60 Operator s Manual This page intentionally left blank ...
Страница 64: ...Resetting Alarms Alarms 6 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 118: ...Service Menu Configuring the TD60 11 10 TMS60 Operator s Manual This page intentionally left blank ...
Страница 194: ...Technical Alarm Messages at the TD60 Troubleshooting 14 12 TMS60 Operator s Manual This page intentionally left blank ...
Страница 216: ...Miscellaneous Accessories 17 6 TMS60 Operator s Manual This page intentionally left blank ...
Страница 235: ...TMS60 Operator s Manual B 1 B EMC ...
Страница 241: ...TMS60 Telemetry Monitoring System Operator s Manual C 1 C FCC Compliance ...
Страница 244: ...FCC Compliance C 4 TMS60 Telemetry Monitoring System Operator s Manual This page intentionally left blank ...
Страница 245: ...TMS60 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3 ...
Страница 249: ......
Страница 250: ...P N 046 007056 00 4 0 ...