7 - 30
BeneVision TM80 Telemetry Monitor Operator’s Manual
7.9.3
Displaying QT/QTc Numerics and Segments
After
QT Analysis
is switched on, QT digital area is displayed on the screen. The location of
the QT digital area is dependent on the display configurations. For how to set the location
of the QT digital area, refer to
1.
Parameter label
2.
QT alarm off symbol
3.
QTc value
4.
Measurement unit
5.
ΔQTc alarm off symbol
6.
ΔQTc value (the difference between the current and baseline QTc values)
7.
QT value
7.9.4 Configuring QT/QTc Alarm Settings
To configure QT/QTc alarm settings:
1.
Tap the QT digital area to enter the
QT/QTc
menu.
2.
Tap
QT/QTc Alarm Setup
.
3.
Select the options described in the following table.
3
1
4
5
6
7
2
Options
Description
Settings*
QTc
Configures whether to trigger
the QTc alarm.
On,
Off
Alarm high limit range:
200 ms to 800ms; the step is 1ms.
The default is
500
for adult and the default
is
480
for pediatric.
Alarm priority: Low,
Med
, High
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...