7 - 16
BeneVision TM80 Telemetry Monitor Operator’s Manual
•
Both the HR value displayed on the telemetry monitor’s main screen and that at
the Central Station can meet relevant standards and are reliable. HR trended,
alarm checking, and arrhythmia are based on the HR calculated at the Central
Station.
•
Please be noted that in case of weak wireless signals, the Central Station faces
the risk of data loss and the HR value calculated by the ECG algorithm for the
Central Station may have certain deviation.
7.6.2 About the HR Digital Area
■
The HR area displays heart rate in the unit of bpm with a resolution of 1 bpm.
■
If the HR measurement is invalid, “- - -” displays in place of the HR value.
■
The HR value displays “0” when the HR value is less than 15 bpm.
7.6.3 ECG Waveform Area
1.
ECG Lead
2.
ECG scale bar
3.
ECG waveform
4.
ECG filter setting
5.
ECG scale
7.6.4 About the ECG Waveform Area
■
The ECG waveform, scale indicator, lead, and filter settings display in the configured
ECG color.
■
The ECG waveform area provides scrolling, real-time waveform data and an erase bar
to provide a time indicator of oldest and new data.
■
The ECG waveform area scrolls the waveform at the configured sweep speed.
■
The ECG waveform area indicates a Pacer indicator when a pace pulse is detected
and Paced is enabled.
1
3
4
5
2
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...