12.11 Auto Enter Monitor Mode
12.11.1Understanding Auto Enter Monitor Mode
Auto Enter Monitor Mode allows the telemetry monitor to switch from telemetry mode to
monitor mode when
No Central Monitoring
or
Lethal Arrhy. Alarm
occurs.In this case,
the display turns from off to on and medical staff should pay more attention to the patient’s
status. If the above alarm disappears, the telemetry monitor switches back to telemetry
mode and the display turns off when the configured
Display Auto Off
time has elapsed.
12.11.2Configuring
Auto Enter Monitor Mode
By default,
Auto Enter Monitor Mode
is set as
No Central Monitoring
.
To configure
Auto Enter Monitor Mode
, follow this procedure:
1.
In the
Maintenance
menu, tap
Auto Enter Monitor Mode
.
2.
Choose the desired option:
◆
No Central Monitoring
: When disconnected from the CMS, the telemetry
monitor switches from telemetry mode to monitor mode and displays no
central monitoring alarm.
◆
Lethal Arrhy. Alarm
: When lethal arrhythmia alarm occurs, the telemetry
monitor switches from telemetry mode to monitor mode and displays lethal
arrhythmia alarm.
◆
Both
No Central Monitoring
and
Lethal Arrhy. Alarm
are selected: When
disconnected from the CMS or lethal arrhythmia occurs, telemetry monitor
turns from telemetry mode to monitor mode and displays corresponding
alarm.
3.
Tap the icon
to exit the
Auto Enter Monitor Mode
menu.
NOTE
•
When the telemetry monitor is connected to the CMS, Auto Enter Monitor mode
can only be configured on the CMS. The configuration is synchronized to the
telemetry monitor.
•
The telemetry monitor switches back to telemetry mode only when all the
abnormal conditions that trigger auto enter monitor mode disappear.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...