8 - 4
BeneVision TM80 Telemetry Monitor Operator’s Manual
■
Press the
key to enter the main menu
→
tap
Parameter Setup
→
tap
Resp
.
8.4.3 Configuring the Resp Setup
To configure the Resp setup, follow this procedure:
1.
In the
Setup
section of the
Resp
menu, select the options described in the following
table.
2.
Tap the
icon to exit the
Resp
menu.
8.4.4 Configuring the Resp Waveform
To configure the Resp waveform, follow this procedure:
1.
In the
Waveform
section of the
Resp
menu, select the options described in the fol-
lowing table.
2.
Tap the
icon to exit the
Resp
menu.
8.4.5 Configuring Resp Alarm Settings
To configure Resp alarm settings, follow this procedure:
1.
In the
Alarms
section of the
Resp
menu, tap
Resp
Alarm Setup
.
2.
Select the options described in the following table.
Options
Description
Settings*
NCM Alarm Delay
The telemetry monitor will trigger
the “No Chest Movement” alarm if
the patient has stopped breathing
for longer than the set apnea time.
10s, 15s,
20s
, 25s, 30s, 35s,
40s
Color
Selects the color of Resp digital data
and waveform.
16 colors
The default color is yellow.
*
The factory default settings are in bold.
Options
Description
Settings*
Gain
Selects the waveform size.
x0.25, x 0.5, x 1,
x2
, x3, x4, x5
Speed
Selects the waveform sweep speed.
3mm/s,
6.25 mm/s
, 12.5
mm/s, 25 mm/s
*
The factory default settings are in bold.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...