A. Introduction

This infrared thermometer is used for measuring 

the temperature of the object's surface, which is 
applicable for various hot, hazardous or hard to 
reach objects without contact safely and quickly. 

This unit consist of Optics, Temperature Sensor 

Signal amplifier, Processing circuit and LCD Display. 
The Optics collected the infrared energy emitted by 
object and focus onto the Sensor. Then the sensor 
translates the energy into an electricity signal. This 
signal will be turned out to be digital shown on the 
LCD after the signal amplifier and processing circuit.

B. Warning & Cautions

1. Warning:  

To avoid the potential situation may cause harm or 
damage to people, please pay attention to the 
following items:
1)

Before you use this unit, check on the plastic 
housing carefully. If there is any damage, do not 
use it

2)

Do not point laser directly at eye or indirectly off 
reflective surfaces.

3)

Do not use this unit in the environment of 
explosive gas, steam or dusty.

2. Caution:

To avoid the damage of the unit or the target, please 
protect from the following situations:
1) EMF (electro-magnetic fields) from arc welders, or 

induction heaters.

2) Thermal shock caused by high temperature changes 

allow 30 minutes for unit to stabilize before use.

3) Do not leave the unit on or near objects of high 

temperature

2) Field of view: Make sure the target is larger than 

the unit's spot size. The smaller the target the 
closer measure distance. When accuracy is 
critical, make sure the ta get is at least twice as 
large as the spot size.

(Page 1)

(Page 2)

(Page 3)

C. Distance to Spot size

1)  When take measurement, pay attention to the 

D stance to Spot Size. As the Distance (D)
from the target surface increases, the spot  size 

(S) of the area measured by the unit becomes 

larger. The Distance to Spot size of the unit is 

12:1

***This unit is equipped with a laser, which is used for 
aiming.

D. Emissivity

Emissivity, Most organic materials and painted or 

oxidized surfaces have an emissivity of 0.95 (pre-set 
in the unit). Inaccurate readings will result from 
measuring shiny or polished metal surfaces. To 
compensate, cover the target surface with masking 
tape or flat black paint. Measure the tape or painted 
surface when the tape or painted reach the same 
temperature as the material underneath.

Material

Emissivity Material

Emissivity

Aluminum

0.3

Iron

0.70

Asbestos

0.95

Lead

0.5

Asphalt

0.95

Limestone

0.98

Basalt

0.70

Oil

0.94

Brass 0.5

Paint

0.93

Brick

0.90

Paper

0.95

Carbon

0.85

Plastic

0.95

Ceramic

0.95

Rubber

0.95

Concrete

0.95

Sand

0.90

Copper

0.95

Skin

0.98

Dirt

0.94

Snow

0.90

Frozen food

0.90

Steel

0.80

Hot food

0.93

Textiles

0.94

Glass(Plate)

0.85

Water

0.93

Ice

0.98

Wood

0..94