Operation and setting
34
5. The test step now appears in the list of test steps.
6. Repeat the procedure to add other test steps to the test sequence (e.g. touch current test direct
test phase reversed # 116).
How to sort the test step sequence
1. Once all test steps required have been added and all test steps not required have been deleted,
the UNIMET® 1100ST can work out a sensible sequence for the test steps. Press
to sort
the test steps.
2. Press F1 to confirm that you wish to sort the test steps.
UNIMET® 1100ST has now sorted the test steps into groups (all visual inspections, all test
steps during which the DUT is not switched on, all test steps during which the DUT is
operated in one phase, all test steps during which the DUT is operated in the other phase,
all function tests, etc.).
A