Commissioning and testing
41
TM800_D00165_00_M_XXEN/06.2015
5.3.2
Settings using the TMK-SET software
Continue with chapter "Tests after parameter setting" on page 42.
Set language
Set password
Set buzzer
Common
acknowledgement on/off
TM800 with BMI8/8 resp.
BMI8/4?
Digital inputs: Set function
and operating mode
Relay: Set function and
operating mode
Set device address
Standard display: enter the
appropriate test
Program alarm addresses
for standard alarm
messages.
Enter system numbers or
system designation
yes
yes
Enter individual alarm
messages for the required
channels
If applicable, use semi-automatic mode.
Enter test addresses
Save parameter settings to
project file and print, if
required
Default setting, send
messages and alarms to
TM800
Adapt the Isometer type to
the test addresses
yes
Use of standard assignment
for medical gases/UPS?
Setting "medical"
Wiring acc. to specifications!
yes
no
Operating messages?
Program operating
messages
yes
Switching commands?
Program switching
commands
yes