Commissioning
Commissioning
iso685-D-B_D00177_05_M_XXEN/07.2017
24
7.2 Initial commissioning
Follow the instructions of the commissioning wizard on the display.
7.2.1 Set language
The language selected here will be used in the menu and for device messages.
7.2.2 Set time and date
Alarm messages in the history memory and the insulation resistance value over time can
only be assigned correctly to the isoGraph when the date and time are set correctly.
Check network function!
When the device is integrated into a network, the influence on the
network has to be checked with the device switched on and off.
Language
5.6.2
o Deutsch
• English
o Espanõl
o Français
o Norsk
Date 5.6.4
14.08.2016
min.
1
max.
31
lle
Commissioning
5.6
Please set the
current date
and time.
2/10
7.2.3 Set system type
By setting the system type the insulation monitoring device can be optimally adapted to
the system to be monitored. The system type is essential information for the insulation
monitoring device in order to determine the insulation resistance correctly.
7.2.4 Set profile
In order to adapt the insulation monitoring device optimally to the system to be monito-
red, select a profile here that suits your system. For an overview of the profiles refer to
. The profile "power circuits" is suitable for most IT systems.
System type
5.6.6
• DC
o AC
o 3AC
lle
Commissioning
5.6
Please choose the
type of system
that best suits
your installation.
5/10
Profile 5.6.8
• Power circuits
o Control circuits
o Generator
o High capacitance
o Inverter>10 Hz
o Inverter<10 Hz
lle
Commissioning
5.6
Please select a
profile that suits
your application.
7/10