Menu mode: Operation and setting
61
MK2430_D00129_02_M_XXEN/09.2019
Possible system number settings:
8.4.3.3
Settings menu 3: Test addresses
Set the bus addresses for insulation monitoring devices (e.g. isoMED427P, 107TD47,
IRDH…), transfer switching and monitoring devices (ATICS®), LIM (Line Isolation
Monitors) and GFCI (Ground Fault Circuit Interrupters) which are required to carry out a
test when the "TEST" button is pressed. The setting can only be made for devices which
have also been activated in the "Alarm addresses" menu and/or programmed for
individual alarm texts. Individual alarm texts are a minimum requirement for:
Channel 1…3
(setting "107TD47")
Channel 1 (setting "IRDHxxx")
1. Exit
Back to the main menu.
2. Edit table
Use the arrow buttons to select the internal BMS bus ad-
dress and confirm your selection by pressing "
↵
".
Use the arrow buttons to select the device type:
LIM
Line Isolation Monitors
GFCI
Ground Fault Circuit Interrupter
MED-Iso
Medical insulation monitoring device
Off
Address deactivated
Press "
↵
" to accept the entry.
Repeat the procedure to activate more addresses.
3. Clear table
Deactivate all addresses ("Off").
nn
Meaning
00
There is no text in line 1 of the alarm message.
01…04
Texts of "System 01" to "System 04" are displayed.
T
Programmed text is displayed.
1.Exit
2.Address: 002
Device type: MED-Iso
---Test addresses----
1.Exit
2.Edit table
3.Clear table
Содержание COMTRAXX MK2430
Страница 14: ...Safety instructions 14 MK2430_D00129_02_M_XXEN 09 2019...
Страница 36: ...Installation and connection 36 MK2430_D00129_02_M_XXEN 09 2019...
Страница 40: ...Commissioning and testing 40 MK2430_D00129_02_M_XXEN 09 2019 5 3 1 Settings on the MK2430...
Страница 41: ...Commissioning and testing 41 MK2430_D00129_02_M_XXEN 09 2019 5 3 2 Settings using the TMK SET software BSV...
Страница 83: ......