Operations Manual 

Item: H2020 and H2024 (-E)   

V1.0 

 

 

 
 

                

 
                                                       

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
I. Specifications: 

Speed:

     

H2020  Fixed 24 RPM  (

20RPM for 230V, 50Hz)

 

H2024  Adjustable from 1 to 70rpm 

Temp. Range:     

A5 to 60C 

Temp. Fluctuation:

  0.9C 

Max Capacity:

    6 x 50ml / 16 x 15ml / 44 x 1.5-2.0ml 

Dimensions:

 (wxdxh) 29.5 x 21.6 x 22.9cm (11.6x8.5x9 in.)  

Weight:

  9lbs / 4kg  

Electrical:

  120V, 60HZ or 230V, 50Hz, 2A 250W   

   

     

 

Operating environment:

  20°C to 45°C  

Warranty:

 2 years 

 

 
 

II. Product Set-Up:

 

Place the rotator  on a clean and  stable surface.  

(Ensure that there is at least 3 inches of open space 

both in front of and behind the instrument.)  Plug 

the instrument into an outlet of the appropriate 

voltage.   

 

III. Tube Holder Installation / Exchange: 

The instrument includes an assortment of tube 

holders and posts.   

 

ALWAYS

  attach the tube holder load tubes outside 

of the incubation chamber.  Once fully loaded, then 

load the posts with tube holders onto the rotating 

axis.  Loading tubes and tube holder within the 
incubation chamber may result in the dropping of 

tubes and spillage on the heating surface. 

 
1.5/2.0, 5, 7, 10 and 15ml Tubes: 

 

 

Attach the tube holders (above) to the included post 
with two arms.  Press holders on firmly to prevent 

them from coming loose during use.  Once pressed 

on firmly, attach the post to the rotator by inserting 

the post into the left side of the rotator first and then 

into the main body (right side).  Once inserted, rotate 

the post until it is fully installed, see below: 
 

 

 

 

 

 

 

 

 

 

To adjust the angle of which the tubes rotate, simply 

rotate the position of the tube holders by hand until 
the desired rotational angle is achieved. 

 

0.5 and 50ml Tubes 

Attach the large tube 

holder directly to the 

rotator (without posts.) 

 

Insert the holder into the 

left side of the rotator first 
and then into the main 

body (right side).  Once 

inserted, rotate the post 

until it is fully installed (similar to diagram above). 

 

To rotate tubes horizontally (rolling/hybridization 

motion), first attach the 2 small posts to the center of 
the large tube holder.  (One post has a small flat that 

can be inserted to correspond to the flat located in 

the center of the tube holder).  Thread the 2 posts 

together by turning the post (without flat) clockwise 

until fully tightener.  Then insert posts into the 

rotator as described above. 
 

To load test-tubes, press them firmly into the 

appropriate holder.  ALWAYS ensure a balanced load 

across both sides of the rotational axis.  (When 

mixing 15 and 50ml tubes, always load a similar 

quantity of tubes on each side of the rotational axis.)  

 

 
 
 
 
 
 
 
 
 
 
 
 
 