- 37 -
Drain Valve
Cleaning of air filter drain valve and water drainage from air compressor
The drain valve is provided for drainage of water from the air filter.
Drain the water accumulated in the air filter by turning the drain valve counterclockwise, at least once
a week.
If water enters the unit interior, failure of air turbine, air motor, syringe and so forth may arise. After
water drainage from the air filter, make sure to close the drain valve by turning it clockwise.
Drain the water accumulated in the air compressor, with the drain valve opened, once a week.
5. CARE AND MAINTANANCE
4. Junction box section
Cleaning of product exterior
Apply dry wiping to metallic components with soft cloth.
If water was spilled to any component, wipe off the moisture immediately. Corrosion and rusting may
arise otherwise.
Wipe the plastic cover with wet and soft cloth.
Sterilization of product exterior
Use FD333 made by Durr for cleaning and bacteria elimination from the product exterior.
NOTICE:
Wipe off water and residual disinfectant immediately. Corrosion and rusting may arise
otherwise.
5. Product exterior section