3. Cuspidor unit section
- 34 -
1
Filter
Bar
Cap
Cleaning of solid collector
Remove the filter in the solid collector located in
the cuspidor unit section on termination of medical
treatment, and wash the filter.
The vacuum suction force will drop if excessive solids
are contained in the solid collector filter.
For removing the filter, remove the cover and pull the
bar. The filter comes out together with the bar. The
filter can be removed from the bar when the filter is
twisted.
Cleaning of cuspidor section
Waste filter in the cuspidor bowl is easily clogged.
Therefore, clean it daily on termination of medical
treatment.
Never use emery paper, metallic brush or detergent containing grinding agent for cleaning.
Use of strong acid detergent or alkali pipe detergent may corrode metals. Never use detergent of
such a nature.
Make sure to turn off the main switch before dismounting or mounting of the cuspidor bowl or
cupfiller tray. If the main switch is not turned off at such an occasion, the cupfiller sensor may be
activated by hand or any other article during mounting/dismounting work, water for bowl flush
may jet out, and corrosion to metallic components may arise.
The cuspidor bowl is made of porcelain, and it may be cracked if impact is applied to it.
Furthermore, cleaning of the cuspidor bowl with hot water should be avoided because cracking
may result.
The cuspidor bowl can be easily removed. Be careful not to apply excessive force at occasions
other than cleaning. (Be careful not to apply impact to it, not to drive it and not to allow it to fall.)
CAUTION
CAUTION
1
5. CARE AND MAINTANANCE