1
Cuspidor Unit Section
2
Flow Adjustment
ADJUSTMENT OF PARTS
Cupfiller Flow Control
• This knob adjusts the water feeding quantity.
• The flow rate decreases when the knob is turned
clockwise, and the flow rate increases when the
knob is turned counterclockwise.
Bowl Flush Flow Control
• This knob adjusts the flow rate of bowl flush.
• The flow rate decreases when the knob is turned
clockwise, and the flow rate increases when the
knob is turned counterclockwise.
Air Vacuum Suction Power Adjustment (Air Vacuum
Specification)
• This knob adjusts the suction power of the air vacuum.
• The suction power decreases when the knob is
turned clockwise and increases when the knob is
turned counterclockwise.
* Flow control knobs are located on the inside of the
cuspidor unit. Loosen the side cover fixing screw by
flat head screwdriver on the rear side of the cuspidor
unit. Open the side cover to outside as shown on the
left figure.
1
Junction Box Section
3
Main air pressure reducing valve
• This knob adjusts the main pressure of air supplied
to the unit.
• Pull the knob and turn as shown in the drawing so
that the main air pressure gauge reads a value
between 0.45 and 0.5 MPa.
• Be sure to push the knob to lock it after setting.
• This knob is adjusted when installing the unit.
Do not change the setting under normal conditions.
Main water pressure reducing valve
• This knob adjusts the main pressure of water
supplied to the unit.
• Pull the knob and turn as shown in the drawing so
that the main water pressure gauge reads a value
between 0.1 and 0.2 MPa.
• Be sure to push the knob to lock it after setting.
• This knob is adjusted when installing the unit.
Do not change the setting under normal conditions.
* The pressure adjustment valves are located on the
inside of the junction box. Loosen the two cover
fixing screws by phillips screwdriver and remove the
junction box cover as shown on the lift figure.
How to open the side cover
Side cover
Side cover
fixing screw
Air vacuum
suction power control
Bowl flush
flow control
Cupfiller
flow control
Flat head
screwdriver
Cupfiller nozzle
Bowl flush nozzle
Bowl flush
nozzle
Cupfiller
nozzle
Main water pressure
reducing valve
Main air pressure
reducing valve
Pull up
Push down
Decrease
Knob
Knob
Pull up
Push down
Increase
Decrease
Increase
How to open the junction box cover
Cover fixing
screw
Phillips screwdriver
Junction box
cover
23
Содержание CREDIA G1
Страница 2: ......