CAUTION
CAUTION
E type:Standard
A type:Option
E
A
E
A
Precautions for water quality safety
To maintain the water quality for treatment and for stable water supply to the handpieces, ensure to carry out
flushout of the residual water in the unit before starting treatment.
At the beginning of the week or in the morning, that is, a long period of idle time, the residual water in tubings
and water heater in the unit tends to permit bacteria being propageted.
Ensure to carry out flushout of water lines in the unit before starting treatment, for the safe treatment and for the
operation of the handpiece without any problem.
It is also recommended to flush out the residual water in the unit and in handpiece tubings with fresh water,
after the treatment in the morning and also in the evening to prevent bacteria propagation.
Standard length of time required for flushout of water lines in the unit
Handpiece line
Air Turbine
Micro Motor
Scaler
Syringe (for both doctor / assistant)
About 40 seconds per each of turbine, motor, scaler and syringe.
(About 40 seconds also in case of flushout of all of them in a batch.)
Apply flushout to the scaler with the body kept fitted. Failure
may result otherwise.
Cuspidor line
Cupfiller (water heater)
Bowl flush
About 5 minutes for cleaning water in cupfiller line.
Flushout procedure
Handpiece line
Pick up one handpiece out of the holder, and hold it
over the cuspidor bowl with the turbine kept as it is
and the contra angle connected to the motor.
(In case the flushout function is provided, pick up
all the handpieces in a lump from the holder, and
hold them on the cuspidor bowl.)
Cuspidor line
Cupfiller (water heater)
Bowl flush
Handpiece line
Pick up one handpiece out of the holder, and hold it
over the cuspidor bowl with the turbine kept as it is
and the contra angle connected to the motor.
( *In case the flushout function is provided, pick up
all the handpieces in a lump from the holder, and
hold them over the cuspidor bowl.)
Cuspidor line
Cupfiller (water heater)
Bowl flush
Flushout procedure
Short time flush out:
For Flushing out the
Handpiece water line only
As for flushout procedures,
please refer to Page 12
selection and operation of
short time flush out.
Long time flush out:
For Flushing out the
cupfiller and cuspidor
bowl after handpiece
flushout
As for flushout procedures,
please refer to Page 12
Selection and operation
of long time flush out.
* Applicable when flush-out function was selected as
optional.
Turn the toggle switch on and flush-out the handpiece
Repeat cupfilling a few times. (at least 3 times)
Press the bowl flush switch and keep flushing the
cuspidor bowl for 5 minutes.
REQUEST FOR WATER FLUSHOUT (DISCHARGE)