  2. Push F switch momentarily until the "a" light above the F switch illuminates. The exposure

            time display window shows the present film speed for "a" setting. (The factory default setting,

            F.09 should be displayed.) By depressing      or       switch, increase or decrease film speed

            number until desired number for "a" setting is displayed

.

  

   

      

  3. To change the "b" setting from the factory default, F.0

4

, push F switch momentarily until the         

            "b" light illuminates. By depressing     or      switch, increase or decrease film speed until the           
             desired number for "b" setting is displayed.

 

      

  4. Press T1 switch to store these settings, then turn the main power switch off. 

     b. Pushing 

Film Speed Selection Switch

 11 momentarily displays the selected film speed setting

 

         

in the 

Exposure Time Display Window 

16

         Depressing this switch for more then 2 sec. changes the film type being selected.

     

c. If the 

Digital Imaging Switch

 12 is depressed, both of the film speed indicating lights (a & b)

         are turned off. 

 

  

 

 

  

 

'LJLWDO,PDJLQJ6ZLWFK

,IDGLJLWDOLPDJLQJV\VWHPLVXVHGVKRUWHUH[SRVXUHWLPHLVRIWHQUHTXLUHG

The 

BELRAY

,,KDV

16 s

SHHGVIRUGLJLWDOLPDJLQJGaG3XVKLQJWKLVVZLWFKPRPHQWDULO\GLVSOD\VWKHVSHHG

EHLQJVHOHFWHGLQWKH

([SRVXUH7LPH'LVSOD\:LQGRZ

:LWKWKHIDFWRU\VSHHGVHWWLQJG

08

WKHH[SRVXUHWLPHEHFRPHVKDOIRI)

08

VHWWLQJ

       As the sensitivity is different according to each manufacturer of digital imaging sensors, this

       setting should be adjusted.  To get a darker image, increase the speed setting and to get a lighter
       image, decrease the speed setting. If the speed setting is increased by 1, exposure time becomes
       12 % longer.
         1. Keep 

4 mA

 selection switch and 

mA selection switch depressed simultaneously for more 

             

than 

3 seconds.  

Release the switches if the ready light starts to flash.

 

         2. Push D switch momentarily until the light above the D switch illuminates and the exposure
             time display window shows the present speed setting. (The factory default setting d.

08

 

             should

 

be displayed.) 

         3. By depressing       or      switch, increase or decrease speed until the desired number is
             displayed. 
         4. Press 

T1 switch

 to store these settings, then turn the main power switch off.

 

 
 
 
 

 
 

 
 
 
  
  

 
 
 
  
 

                                                                                  

 

 

 13

4 mA

 S

election 

S

witch

     

By m

omentarily depressing this switch

, the tube current is set at 4 mA

     

When Film switch is depressed, the tube current setting will be automatically changed to 7 mA.

 

 14

  

7 mA

 S

election 

S

witch

     

By mo

mentarily depressing this switch

, the tube current is set at 7 mA.

    

 When digital switch is depressed, the tube current setting will be automatically changed to 4 mA.

 

5

  

Содержание BELRAY II 097

Страница 1: ...R OPERATOR S INSTRUCTIONS for USA Canada WARNING This X ray unit may be dangerous to patient and operator unless safe exposure factors operating instructions and maintenance schedules are observed DEN...

Страница 2: ...F CONTROLS 4 4 OPERATING PROCEDURES 7 5 ESTIMATED AIR KERMA 7 6 OPTIONAL HAND EXPOSURE SWITCH 7 7 DIGITAL IMAGING SYSTEM 8 8 DISINFECTION AND CLEANING 8 9 ERROR CODES 9 10 MAINTENANCE 10 11 DISPOSAL 1...

Страница 3: ...housing assembly 097 H b X ray controls 097 CM main controller 097 CS sub controller c Cones 097 R regular 097 L long d Balance arm 097 A 4 COMPLIANCE WITH STANDARD The BELMONT BELRAY II MODEL 097 x r...

Страница 4: ...re warning light be at least 2 m away from the x ray head and patient and out of the path of the x ray beam or be positioned behind a protective device fully use all radiation protection devices acces...

Страница 5: ...justing Switch Down 12 Digital Imaging Switch 4 ExposureTimeAdjusting Switch Up 13 4 mA Selction Switch 5 Tooth Selection Switch T1 6 Tooth Selection Switch T2 7 Tooth Selection Switch T3 8 Tooth Sele...

Страница 6: ...I 0DQGLEOH 7 QFLVRU RI 0D LOOD XVSLG 3UHPRODU RI 0DQGLEOH 7 XVSLG 3UHPRODU RI 0D LOOD 0RODUV RI 0DQGLEOH LWHZLQJ 7 0RODU RI 0D LOOD LWHZLQJ 0RODUV 7 2FFOXVDO I WKH 7 VZLWFK LV GHSUHVVHG PRUH WKDQ VHF...

Страница 7: ...PRPHQWDULO GLVSOD V WKH VSHHG EHLQJ VHOHFWHG LQ WKH SRVXUH 7LPH LVSOD LQGRZ LWK WKH IDFWRU VSHHG VHWWLQJ G 08 WKH H SRVXUH WLPH EHFRPHV KDOI RI 08 VHWWLQJ As the sensitivity is different according to...

Страница 8: ...osure and an error code E 00 will be displayed in Exposure Time Display Window 16 6 switch once after setting to child If the weight of child is over 30kg and less than 50kg press Speed Setting mA Chi...

Страница 9: ...6ZLWFK in RUGHU WR SUHYHQW DFFLGHQWDO exposures when the unit is not in XVH 127 I WKH XQLW OHIW RYHU PLQ ZLWKRXW EHLQJ RSHUDWHG DQG WKH 0DLQ 3RZHU 6ZLWFK LV NHSW RQ ILJXUH UXQV WKURXJK WKH SRVXUH 7LPH...

Страница 10: ...QLQJ RI WKH SDUWV WKDW PD FRPH LQ FRQWDFW ZLWK VNLQ SHULRGLF GLVLQIHFWLRQ ZLWK D QRQ FRUURVLYH VXUIDFH GLVLQIHFWDQW LV UHFRPPHQGHG 5HFRPPHQGHG GLVLQIHFWDQW XUU HQWDO 237 0 7 6FL DQ WG U LQJ OORZ VXUID...

Страница 11: ...Malfunction of the microcomputer Tube current is detected during pre heating period Tube current is detected when main power switch is turned on Exposure switch or exposure circuit had been ON when ma...

Страница 12: ...ed service technician The specific instructions to perform these checks are located within the BELRAY II MODEL 097 Installation Manual A Line voltage confirmation B Tube current confirmation C Inspect...

Страница 13: ...long cone 1 9 3 3 mGy sec 40 Data obtained by direct measurement in the useful beam 16 Nominal electrical output of H V Generator 0 36kW at 70kVp 7mA 17 Cone Source to skin distance Field size a Regu...

Страница 14: ...netic emissions Emissions test Test procedure Compliance Note The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals CISPR 11 class A If it is used...

Страница 15: ...ding cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Radiated RF IEC61000 4 3 80 MHz 2700 MHz 3V m unmodulated r m s 80 AM 1kHz 80 MHz 2700...

Страница 16: ...w Jersey 08873 U S A TEL 732 469 5000 800 223 1192 Fax 732 526 6322 800 280 7504 www belmontequip com TAKARA COMPANY CANADA LTD 2455 Meadowvale Boulevard Mississauga Ontario L5N 5S2 Canada TEL 905 816...

Отзывы: