Step-By-Step Quick Set-Up

 

SETUP:

 

 

Inspect the system

 

  Power cord: 14/3 SJT cable with Hospital plug 

(

USE ONLY SUPPLIED POWER CORD

)

 

  IV pole: 5 wheel IV pole (max. diameter 1 1/4") 

  Pole mount support assembly (Support Clamp & Plastic washer) 

  Reservoir support 

  Disposable set 

  3L Reservoir and reservoir holder, if needed 

 

 

 

MOUNT ON THE IV POLE:

 

30 in (76 cm)

Pole Clamp

Release Handle

 

SETUP / MOUNT ON IV POLE

 

  Mount The Belmont

® 

Rapid Infuser 

RI-2

app. 30” from the ground 

  Tighten the “Pole Clamp Release Handle” 

onto the IV pole 

  Install the Pole Mount Support Assembly 

(support clamp and plastic washer) under 

the 

RI-2

 

  Clamp the Reservoir Support above the 

RI-2

 

Содержание RAPID INFUSER RI-2

Страница 1: ...pport assembly Support Clamp Plastic washer Reservoir support Disposable set 3L Reservoir and reservoir holder if needed MOUNT ON THE IV POLE 30 in 76 cm Pole Clamp Release Handle SETUP MOUNT ON IV POLE Mount The Belmont Rapid Infuser RI 2 app 30 from the ground Tighten the Pole Clamp Release Handle onto the IV pole Install the Pole Mount Support Assembly support clamp and plastic washer under the...

Страница 2: ...fusing blood components and solutions compatible with blood Using aseptic technique make patient connection without entrapping air Press INFUSE Infusion starts at 10 ml min Press INFUSE RATE to change flow rate BATTERY System automatically switches to battery when AC is disconnected BOLUS INFUSION Infuse fixed volume at 200 ml min Fixed volume Factory set at 200 ml Return to previous flow rate if ...

Страница 3: ...lly opened Bag not fully spiked Tubing in Fluid OUT Sensor not fully seated Filter clogged Recirculate line smaller tubing blocked Add fluid Open bag clamps Check bag spikes Reseat tubing in Fluid OUT Sensor Replace filter if it becomes clogged Check recirculate line HIGH PRESSURE Infusion site may be positional Cannula bore size is too small Patient line is blocked Recirculate line is kinked Pres...

Страница 4: ...or errors Power off and restart Service machine if error persists SYSTEM ERROR 203 Excessive AC power line noise or internal failure Press RETRY to continue Service machine if error persists SYSTEM ERROR 204 Power feedback circuit errors Power off and restart Service machine if error persists SYSTEM ERROR 205 Heater hardware errors Press RETRY to continue Service machine if error persists SYSTEM E...

Страница 5: ...SPONDING Key pad sensitivity has been set to SLOW Key pad errors To reset key pad sensitivity see SYSTEM PARAMETERS SETTING below Power down and retry If error persists call service WILL NOT HEAT Windows on disposable set or IR probes are wet dirty or obstructed Power Driver module is not calibrated properly IR probes are out of calibration Clean IR probes with moistened Q tip and dry Replace disp...

Страница 6: ... Factory set at level 4 KEY RATE Press KEYRATE key There are three 3 levels of key pad sensitivity FAST MEDIUM SLOW Factory set at MEDIUM Press to select the time required to depress a key for stoke to be recognized SLOW requires the most time and makes the touch key less sensitive BOLUS VOLUME Press SETUP BOLUS key Factory set at 200 ml Change the preset volume Press and hold BOLUS key The bolus ...

Страница 7: ...ct the larger pump tubing by pressing in the luer lock tab and pulling out the connector Disconnect the thinner recirculate line by unscrewing the connector 3 Connect the reservoir chamber to the luer connectors of the 3 Spike disposable set INSTALL 3 LITER RESERVOIR 1 Assemble the 3L Reservoir using aseptic techniques by attaching the three fluid supply tails onto the top of the reservoir 2 Attac...

Страница 8: ... IR Probe Probe IR 780 BOSTON ROAD BILLERICA MA 01821 USA USA 866 663 0212 WORLDWIDE 1 978 663 0212 WWW BELMONTMEDTECH COM 702 00201 Rev D CONVERSIONS HEATING Heats to 39o C Heats to 37 5o C ml min ml hr ml min ml hr 2 5 150 60 3 600 5 300 70 4 200 10 600 80 4 800 20 1 200 90 5 400 30 1 800 100 6 000 40 2 400 200 12 000 50 3 000 300 18 000 400 24 000 500 30 000 600 36 000 700 42 000 750 45 000 100...

Страница 9: ...UIDE TECHNICAL SUPPORT USA 855 397 4547 WORLDWIDE 1 978 663 0212 TECHSUPPORT BELMONTMEDTECH COM WWW BELMONTMEDTECH COM IT IS ESSENTIAL THAT YOU READ AND UNDERSTAND THE OPERATOR S MANUAL BEFORE USING THIS DEVICE Registered in Accordance with ISO 13485 ...